LUMIRADX SARS-COV-2 AG TEST STRIP KIT
Report
- Report Number
- 3012642695-2021-01270
- Event Type
- Malfunction
- Date Received
- July 30, 2021
- Date of Event
- July 5, 2021
- Report Date
- July 30, 2021
- Manufacturer
- LUMIRADX UK LTD.
- Product Code
- QKP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- PHYSICIAN
Narratives
PATIENT WAS REPORTED TO BE SYMPTOMATIC ("PATIENT HAD A COLD") WHEN INITIAL LUMIRADX SARS-COV-2 AG TEST WAS PERFORMED ON (B)(6) 2021 WITH STRIP LOT 6000256 AND INSTRUMENT SERIAL NUMBER (B)(4). SECONDARY TESTING WAS PERFORMED ON AN ALTERNATIVE INSTRUMENT OF ANOTHER MANUFACTURER'S MAKE AND MODEL (AEGLE CORONAVIRUS AG RAPID TEST) WHICH PRODUCED A NEGATIVE RESULT. THE CUSTOMER REPORTED THE FOLLOWING PROCESS FOR TESTING ON THE LUMIRADX PLATFORM: "SAMPLE IS TAKEN AND MEASURED DIRECTLY." THE CUSTOMER REPORTED THEIR CLEANING PRACTICES TO BE "STRIPS ARE REMOVED DIRECTLY AND THE INSTRUMENT IS WIPED WITH A DRY CLOTH", AND FURTHER REPORTED THAT GLOVES ARE CHANGED AFTER EACH SAMPLE. LOT 6000256 MET ALL DEFINED QC CRITERIA AT THE TIME IT WAS RELEASED AND TESTING USING NEGATIVE NASAL SWABS FROM IN-HOUSE DONORS MEETS EXPECTED PERFORMANCE FOR USE IN THE FIELD. REVIEW OF PRODUCT RISK ASSESSMENT - SARS-COV-2 AG ASSAY REVISION 7, RESULTED IN SEVERITY OF MODERATE, AS FOLLOWS: DELAYED DIAGNOSIS LEADING TO DELAYED TREATMENT OF THE TRUE CAUSE OF THE PATIENT'S SYMPTOMS. UNNECESSARY ISOLATION. NO PATIENT HARM, INJURY OR ADVERSE HEALTH CONSEQUENCES WERE COMMUNICATED BY THE CUSTOMER TO LUMIRADX FOR THE REPORTED DISCORDANT RESULT. TRENDING DATA FOR DISCORDANT RESULTS WAS REVIEWED FOR THIS LOT AND THE OCCURRENCE RATE PER QUANTITY OF STRIPS IN THE FIELD WAS CALCULATED AS (B)(4). LUMIRADX SARS-COV-2 AG TEST PRODUCT INSERT CLAIMS A SPECIFICITY OF 96.6% WITH A REFERENCE RT-PCR ASSAY AND IT IS ACCEPTED THAT UP TO 3.4% OF TEST STRIPS MAY GENERATE A DISCORDANT FALSE POSITIVE RESULT. LUMIRADX R&D REVIEW OF COMPLAINT DATA IDENTIFIED THAT AEGLE CORONAVIRUS AG RAPID TEST IS A LATERAL FLOW TEST, WITH A LOWER SENSITIVITY THAN THE LUMIRADX SARS-COV-2 AG TEST - AEGLE LOD IS APPROXIMATELY 4-FOLD LOWER (115 TCID50/ML VS 32 TCID50/ML), THAN LUMIRADX; THEREFORE THE SIMILARITY IN CLINICAL AGREEMENT MOST LIKELY COMES FROM AEGLE SAMPLE SET CONTAINING A HIGHER VIRAL LOAD. THE REDUCED SENSITIVITY AND DIFFERENCES IN LOD, SPECIFICALLY AT LOW VIRAL TITERS, SUGGESTS THAT IT IS ENTIRELY POSSIBLE THESE RESULTS REPRESENT A TRUE POSITIVE FROM LUMIRADX TEST AND A FALSE NEGATIVE FROM THE AEGLE TEST. ROOT CAUSE DETERMINATION: NO DEFINITIVE ROOT CAUSE HAS BEEN DETERMINED FOR THE REPORTED DISCORDANT RESULT BASED ON THE INFORMATION CURRENTLY AVAILABLE. INVESTIGATION CONCLUSION AND STATUS OF INVESTIGATION: NO FURTHER INVESTIGATION IS CONSIDERED NECESSARY AT THIS TIME. THIS STRIP LOT CONTINUES TO DEMONSTRATE FIELD PERFORMANCE WITHIN SPECIFICATION OF PRODUCT CLAIMS RELATIVE TO THE QUANTITY OF STRIPS FROM THIS LOT IN THE FIELD. REPORTS OF DISCORDANT RESULTS FOR THIS LOT WILL CONTINUE TO BE TRENDED AND REVIEWED, WITH ACTION TAKEN AS APPROPRIATE IN RESPONSE TO ANY ADVERSE TRENDS OR EVENTS INCLUDING A FOLLOW-UP REPORT.
THIS IS REPORT 2 OF 3 FOR THIS FACILITY AND AWARE DATE. THE CUSTOMER REPORTED A SUSPECTED FALSE (INCORRECT) POSITIVE RESULT FROM A RAPID RESULT COVID TEST SYSTEM ON AN INDIVIDUAL SYMPTOMATIC PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1148855 | LUMIRADX SARS-COV-2 AG TEST STRIP KIT | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM | QKP | LUMIRADX UK LTD. | 6000256 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |