FDA Adverse Event Malfunction Summary report: N

DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS

MDR report key: 12253261 · Received July 30, 2021

Report

Report Number
2939274-2021-04352
Event Type
Malfunction
Date Received
July 30, 2021
Report Date
July 1, 2021
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTJ
UDI-DI
10886982189943
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6: PART #: 319.006, SYNTHES LOT #: 7961855, SUPPLIER LOT #: N/A, RELEASE TO WAREHOUSE DATE: 08 APR 2015, MANUFACTURED BY: JENNERSVILLE, NO NCR'S GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. VISUAL INSPECTION: THE DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS (P/N: 319.006, LOT #: 7961855) WAS RETURNED AND RECEIVED AT US CUSTOMER QUALITY (CQ). UPON VISUAL INSPECTION, IT WAS OBSERVED THAT THE NEEDLE COMPONENT WAS BROKEN. ADDITIONALLY, SCRATCHES FROM FIELD USAGE WERE OBSERVED ON THE DEVICE. NO OTHER ISSUES WERE IDENTIFIED. THE PROVIDED PHOTOGRAPH WAS REVIEWED AND NO ADDITIONAL ISSUES WERE OBSERVED. DEVICE FAILURE/DEFECT IDENTIFIED? YES. DIMENSIONAL INSPECTION: DIMENSIONAL ANALYSIS CANNOT BE PERFORMED DUE TO POST-MANUFACTURING DAMAGE. DOCUMENT/SPECIFICATION REVIEW: THE FOLLOWING DOCUMENTS WERE REVIEWED: CURRENT AND MANUFACTURED NO DESIGN ISSUES OR DISCREPANCIES WERE IDENTIFIED. COMPLAINT CONFIRMED? YES. INVESTIGATION CONCLUSION THE COMPLAINT CONDITION WAS CONFIRMED FOR THE RETURNED DEVICE. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED BASED ON THE PROVIDED INFORMATION. THE UNINTENDED EXTERNAL FORCES MIGHT HAVE CONTRIBUTED TO THE REPORTED COMPLAINT CONDITION. NO NEW, UNIQUE OR DIFFERENT PATIENT HARMS WERE IDENTIFIED AS A RESULT OF THIS EVALUATION. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. NO DESIGN ISSUES WERE OBSERVED DURING THE DOCUMENT/SPECIFICATION REVIEW. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

REPORTER IS A J&J SALES REPRESENTATIVE. THE SUBJECT DEVICE HAS BEEN RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON AN UNKNOWN DATE, THREE (3) DEPTH GAUGES FOR 2.0MM AND 2.4MM CORTEX SCREWS WERE BROKEN. IT IS UNKNOWN IF THERE WAS PATIENT AND PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE (1) DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1150902 DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS GAUGE, DEPTH HTJ WRIGHTS LANE SYNTHES USA PRODUCTS LLC 319.006 7961855 10886982189943

Patients

Seq Age Sex Outcome Treatment
1