MONOPLUS VIOLET 3/0 (2) 70CM HR22 (M)RCP
Report
- Report Number
- 3003639970-2021-00328
- Event Type
- Malfunction
- Date Received
- July 30, 2021
- Date of Event
- July 8, 2021
- Report Date
- July 30, 2021
- Manufacturer
- B. BRAUN SURGICAL, S.A.
- Product Code
- NEW
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S., K031216. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE BATCH OF WHICH WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE BATCH. THERE ARE NO UNITS IN STOCK IN B. BRAUN SURGICAL'S WAREHOUSE. WE HAVE 1 OPEN SAMPLE CONTAINING ONLY THE NEEDLE, THE THREAD IS MISSING. WITHOUT CLOSED SAMPLES A PROPER ANALYSIS CANNOT BE PERFORMED. REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFILL USP/EP AND B. BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: IN SPITE OF RECEIVING A DEFECTIVE SAMPLE, WITHOUT CLOSED SAMPLES A SUITABLE ANALYSIS CANNOT BE PERFORMED. NEVERTHELESS, WE TAKE NOTE OF THIS INCIDENCE AND IF ANY CLOSED SAMPLE IS RECEIVED IN THE FUTURE, WE WILL RE-OPEN THE CASE AND ANALYSE IT. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.
IT WAS REPORTED AN ISSUE WITH MONOSYN SUTURE. THE CLIENT REPORTED THAT WHEN OPENING THE PACKAGE AND GRASPED THE SUTURE, THE NEEDLE CAME OFF. THERE IS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1151859 | MONOPLUS VIOLET 3/0 (2) 70CM HR22 (M)RCP | SYNTHETIC ABSORBABLE MONOFILAR SUTURE | NEW | B. BRAUN SURGICAL, S.A. | C2024135 | 120232 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |