IMP TM 3.7MM MTX FULL,11.5MM
Report
- Report Number
- 0002023141-2021-01961
- Event Type
- Injury
- Date Received
- July 29, 2021
- Date of Event
- May 15, 2021
- Report Date
- January 26, 2022
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019003
- PMA / PMN Number
- K132258
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- 003
Narratives
MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2021-01960-1. THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. D9: DEVICE AVAILABILITY WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODE WAS ADDED: 3331 AND 4114. H10: NARRATIVE/DATA WAS UPDATED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBERS: (1231751 & 1232677). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WERE NOTED AS PART OF THE DHR. LOTS WERE INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBERS: (1231751 & 1232677) FOR SIMILAR EVENTS (COMPLAINT CATEGORY KEYWORD: FUNCTIONAL: UNABLE TO PLACE IMPLANT INTO OSTEOTOMY) AND 3 OTHER COMPLAINTS WERE IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE.
MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2021-01960-2. THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: NO UPDATE TO THE INVESTIGATION RESULTS. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE.
NO ADDITIONAL EVENT INFORMATION AT THE TIME OF THIS REPORT.
NO ADDITIONAL EVENT INFORMATION AT THE TIME OF THIS REPORT.
MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT. PLEASE SEE ASSOCIATED REPORT: 0002023141-2021-01960. ZIMMER BIOMET COMPLAINT NUMBER (B)(4). A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR LACK OF PRIMARY STABILITY EVENTS THAT DO NOT ALLEGE A DEFICIENCY WITH THE IMPLANT AND IDENTIFIED THAT THE REPORTED EVENT IS LIKELY DUE TO BIOLOGICAL FACTORS WHICH HAVE AN ADVERSE EFFECT ON IMPLANT STABILITY OR IS RELATED TO SURGICAL TECHNIQUE AND INSERTION TORQUE. DUE TO A WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED), IDENTIFYING A DEFINITIVE ROOT CAUSE IS GENERALLY NOT POSSIBLE. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. 3
DOCTOR INDICATED LACK OF PRIMARY STABILITY. PATIENT TO RESCHEDULE. TOOTH SITES 45-23.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1144819 | IMP TM 3.7MM MTX FULL,11.5MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TMTB11 | 1232677 | 00889024019003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | Required Intervention |