FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM

MDR report key: 12244735 · Received July 29, 2021

Report

Report Number
0002023141-2021-01952
Event Type
Injury
Date Received
July 29, 2021
Date of Event
March 8, 2021
Report Date
December 17, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019508
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. B5: DESCRIBE EVENT OR PROBLEM. D9: DEVICE AVAILABLE FOR EVALUATION CHANGE ¿NO' TO 'YES'. G3: DATE RECEIVED BY THE MANUFACTURER. H1: TYPE OF REPORT, FOLLOW-UP NUMBER. H2: FOLLOW UP TYPE. H6: EVALUATION CODES. H10: ADDITIONAL NARRATIVE DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1230358). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1230358) FOR SIMILAR EVENTS (KEYWORD CATEGORY: DENTAL: MEDICAL: BONE LOSS) AND NO OTHER COMPLAINT WAS IDENTIFIED

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED, PMA/510(K) NUMBER K011028 AND K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCE AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THE IMPLANT WAS REMOVED DUE TO BONE LOSS. 5 MONTHS LATER THE PATIENT BACK FOR A REVIEW WITH SOME PAIN, THE TEST SHOWED THAT THERE WAS BONE LOSS, THE IMPLANT WAS REMOVED AND A NEW IMPLANT WAS IMPLANTED THE SAME DAY WITH A BIOS BONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1146236 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVB10 1230358 00889024019508

Patients

Seq Age Sex Outcome Treatment
1 71 YR Female Required Intervention