PN RELION 31G X 6MM
Report
- Report Number
- 3014704491-2021-00051
- Event Type
- Malfunction
- Date Received
- July 28, 2021
- Date of Event
- June 29, 2021
- Report Date
- July 29, 2021
- Manufacturer
- BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
- Product Code
- FMI
- UDI-DI
- 80382903206175
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H6: INVESTIGATION: CUSTOMER RETURNED (4) SEALED 6MM, 31G PEN NEEDLES FROM LOT # 0140337. CUSTOMER STATES THAT THE INSULIN DOES NOT FLOW WHEN PRIMING. ALL RETURNED PEN NEEDLES WERE TESTED AND ALL WERE ABLE TO EXPEL PROPERLY WITHOUT ANY OBSERVED DEFECTS. LOT#0140337 DHR WAS REVIEWED AND NO QN FOUND. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED.
IT WAS REPORTED THAT THE PN RELION 31G X 6MM WAS UNABLE TO DELIVER INSULIN, AND UNABLE TO PRIME. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RELION PHARMACIST AND CONSUMER CALLED IN. CONSUMER REPORTED, NO INSULIN FLOW WHEN PRIMING. STATED, HE PRIMES 2 TO 3 UNITS. STATED, NO INSULIN FLOW WHEN TAKING INJECTION. STATED, HE'S USING 2 OR 3 PEN NEEDLES TO COMPLETE ONE INJECTION.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE PN RELION 31G X 6MM WAS UNABLE TO DELIVER INSULIN, AND UNABLE TO PRIME. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RELION PHARMACIST AND CONSUMER CALLED IN. CONSUMER REPORTED, NO INSULIN FLOW WHEN PRIMING. STATED, HE PRIMES 2 TO 3 UNITS STATED, NO INSULIN FLOW WHEN TAKING INJECTION. STATED, HE'S USING 2 OR 3 PEN NEEDLES TO COMPLETE ONE INJECTION.