SYRINGE 1ML LS SP120
Report
- Report Number
- 3003152976-2021-00432
- Event Type
- Malfunction
- Date Received
- July 28, 2021
- Date of Event
- June 29, 2021
- Report Date
- September 13, 2021
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: TWENTY TWO SAMPLES AND PHOTOS RECEIVED FOR INVESTIGATION, UPON VISUAL INSPECTION OF THE SAMPLES RECEIVED, NO MOLDING OR OTHER DEFECT CAN BE NOTICED IN THE SYRINGE, NO BURRS OR DEFORMATIONS CAN BE SEEN IN THE TIP. ADDITIONALLY, TEN RETAINED SAMPLES OF LOT 200612 WERE USED TO CONDUCT A LEAKAGE TEST. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, AND NO LEAK WAS IDENTIFIED. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2006012, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME.
IT WAS REPORTED THAT THE SYRINGE 1ML LS SP120 WAS DIFFICULT TO CONNECT TO THE TFK LABORATORY NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: "THIS IS A REPORT ABOUT SYRINGE NEEDLE CONNECTIVITY ISSUE. THE CUSTOMER USES THIS PRODUCT FOR COVID VACCINATION." "THE CONNECTION OF THIS SYRINGE (CAT#303172) TO TSK LABORATORY¿S NEEDLE BECAME LOOSE IN 20 OF 320 PRODUCTS."
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE SYRINGE 1ML LS SP120 WAS DIFFICULT TO CONNECT TO THE TFK LABORATORY NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "THIS IS A REPORT ABOUT SYRINGE NEEDLE CONNECTIVITY ISSUE. THE CUSTOMER USES THIS PRODUCT FOR COVID VACCINATION." "THE CONNECTION OF THIS SYRINGE (CAT#303172) TO TSK LABORATORY¿S NEEDLE BECAME LOOSE IN 20 OF 320 PRODUCTS."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1137110 | SYRINGE 1ML LS SP120 | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2006012 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |