FDA Adverse Event Malfunction Summary report: N

CHLORAPREP ONE STEP HI-LITE ORANGE

MDR report key: 12240139 · Received July 28, 2021

Report

Report Number
3004932373-2021-00345
Event Type
Malfunction
Date Received
July 28, 2021
Date of Event
March 12, 2021
Report Date
July 23, 2021
Manufacturer
CAREFUSION 213, LLC 0113
Product Code
KXG
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4). SAMPLES WERE PROVIDED FOR EVALUATION. VISUAL EXAMINATION OF THE SAMPLES SHOWS BLACK SPECS ON SOME OF THE SAMPLES. THE SPECS WERE DIFFICULT TO FIND DUE TO THERE SMALL DIMENSIONAL SIZE. AS A RESULT, THE FAILURE MODE OF BLACK DOTS WAS VERIFIED ON THE SAMPLES ANALYZED. ONE POSSIBLE ROOT CAUSE OF THE BLACK SPECS IS THE PRINTER OF THE PACKAGING LIDDING. PRODUCTION RECORD REVIEW WAS COMPLETED FOR BATCHES/LOTS 1054667, 1033703, 1055787 AND NO NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING OF THIS LOT. ON JUNE 30, 2021, THE MANUFACTURER UPDATED CURRENT WEEKLY PREVENTIVE MAINTENANCE TO INCLUDE A THOROUGH CLEANING OF ALL ROLLERS (ADDED IDLE ROLLERS ON THE PRINTER MACHINE, MARK ANDY EQUIPMENT). NO FURTHER ACTIONS ARE REQUIRED AT THIS TIME. THIS FAILURE MODE WILL CONTINUE TO BE TRACKED AND TRENDED.

Description of Event or Problem · 1

IT WAS REPORTED PACKAGING ISSUES. IT WAS CLARIFIED THAT BLACK DOTS APPEAR AND SMALL SPOTS OF CHLORAPREP LEAKING OUT PRIOR TO OPENING PER RESPONSE EMAIL: IT IS THE SAME ISSUE. THEY ARE HAVING BLACK DOTS APPEAR AND AND SMALL SPOTS OF CHLORAPREP LEAKING OUT PRIOR TO OPENING. I HAVE SPOKEN TO THE CUSTOMER MANY TIMES ABOUT THIS PER EMAIL: OF THE LAST 7 BX'S OF BD# 930815 WE'VE RECEIVED, WE HAVE HAD TO REMOVE 110 EA. FROM STOCK FOR THE SAME PACKAGING ISSUES WE WERE PREVIOUSLY HAVING. BREAKDOWN BY LOT NUMBER BELOW. IS BD GOING TO LOOK INTO THIS ISSUE? CAN YOU ISSUE RESTOCK FOR THE DAMAGED PRODUCT?

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1134822 CHLORAPREP ONE STEP HI-LITE ORANGE 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL KXG CAREFUSION 213, LLC 0113 1055787

Patients

Seq Age Sex Outcome Treatment
1 Other