FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 1223734 · Received November 10, 2008

Report

Report Number
2017865-2008-03523
Event Type
Injury
Date Received
November 10, 2008
Date of Event
September 7, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
PMAP95022
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY, DUE TO OVERSENSING ON THE VENTRICULAR LEAD. A CHEST X-RAY REVEALED THAT THE LEAD HAD DISLODGED TOWARDS THE ATRIUM. THE LEAD WAS REPOSITIONED AND REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120/60 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention