PERCOR STAT DL 9.5 FR 40 CC IAB
Report
- Report Number
- 2248146-1997-01014
- Event Type
- Malfunction
- Date Received
- September 24, 1997
- Date of Event
- September 11, 1997
- Report Date
- September 11, 1997
- Manufacturer
- DATASCOPE CORP.
- Product Code
- DSP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NURSE
Narratives
DEVICE LABEL CODE FOR F10. POSITION 1: 1701. POSITION 2: 1738. EVALUATION: THE IAB WAS RECEIVED INTACT FOR EVALUATION WITH A MODERATE AMOUNT OF BLOOD NOTED TO BE WITHIN THE MEMBRANE AND CATHETER TUBING. THE SHEATH TUBING WAS NOTED TO BE SEVERELY KINKED ALONG ITS LENGTH. NO LEAK WAS FOUND IN THE INNER LUMEN. UNDERWATER LEAK TESTING REVEALED NUMEROUS LEAKS IN THE MEMBRANE. VISUAL EXAMINATION REVEALED SMOOTH-EDGED TEARS IN THE MEMBRANE. THE PRIMARY BALLOON PENETRATION COULD NOT BE LOCATED. PROBABLE CAUSE OF DIFFICULTY: BASED ON THE POOR CONDITION OF THE RETURNED IAB, IT WAS NOT POSSIBLE TO LOCATE THE PRIMARY PENETRATION WHICH ALLOWED BLOOD TO ENTER THE DEVICE. IT IS POSSIBLE THAT THE DAMAGE TO THE BALLOON MEMBRANE OBSCURED THE PRIMARY PENETRATION.
EVENT: (CC# 97-01032) AFTER IABP FOR ABOUT TWO HOURS, "CHECK IAB CATH" ALARM SOUNDED FROM THE PUMP AND THE IAB LEAKED AND WAS REMOVED FROM THE PT. THE FOLLOWING WAS REPORTED TO DATASCOPE ON 10/3/1997: THE PT HAD CABG ON 9/11/1997. FOLLOWING CABG, THE PT WAS HEMODYNAMICALLY UNSTABLE. THE IAB WAS INSERTED IN THE O.R. AT APPROXIMATELY 12:30 PM. TWO HOURS LATER, THE CATHETER LEAKED AND WAS REMOVED AT 3:30 PM ON 9/11/1997. THE PT REMAINED IN CRITICAL CONDITION. (ON 6/17/1998, DATASCOPE RECEIVED THE VOLUNTARY MEDWATCH FROM THE FDA; MDR ACCESS NUMBER 1013816) [EVENT COMPLICATIONS]: UNKNOWN - REPORTED 9/11/1997; NONE - REPORTED 10/3/1997. [PT'S CURRENT STATUS]: STABLE - RPT'D 10/3/1997.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCOR STAT DL 9.5 FR 40 CC IAB | INTRA-AORTIC BALLOON CATHETER | DSP | DATASCOPE CORP. | 0684-00-0283 | 08/01/99 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |