FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION LEAD

MDR report key: 1223086 · Received November 10, 2008

Report

Report Number
2017865-2008-03765
Event Type
Injury
Date Received
November 10, 2008
Date of Event
August 22, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
PMAP95022
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

ANALYSIS REVEALED THAT THE IS-1 PROXIMAL COIL WAS FRACTURED. THE LOCATION OF THE DAMAGE IS CONSISTENT WITH THAT CAUSED BY FRICTION WITH THE ICD CAN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPIES. PER PATIENT REQUEST, BOTH ICD AND LEAD WERE EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ACTIVE FIXATION LEAD DEFIBRILLATION LEAD LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention