FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 1223024 · Received November 10, 2008

Report

Report Number
2017865-2008-03238
Event Type
Injury
Date Received
November 10, 2008
Date of Event
August 13, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DTB
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NA

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED WITH 2:1 BLOCK WITH INTERMITTENT VENTRICULAR CAPTURE AT OUTPUTS LOWER THAN 7 V AT 1.0 MS. LEAD IMPEDANCES IN BOTH UNIPOLAR AND BIPOLAR WERE 244 AND 303 RESPECTIVELY. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENDRIL ST PERMANENT PACEMAKER ELECTRODE DTB ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1788T/52 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention