FDA Adverse Event Malfunction Summary report: N

QUICKIE LXI

MDR report key: 12229945 · Received July 27, 2021

Report

Report Number
9616084-2021-00017
Event Type
Malfunction
Date Received
July 27, 2021
Date of Event
May 18, 2018
Report Date
May 18, 2018
Manufacturer
SUNRISE MEDICAL PRIVADA MISIONES
Product Code
IOR
PMA / PMN Number
K123975
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BACKGROUND INFORMATION: IT IS UNKNOWN THE CONDITION OF THE WHEELCHAIR AT THE TIME OF THE INCIDENT. IN THE LX/LXI OWNER'S MANUAL, PART NUMBER MK-100066 REV. B, IT IS OUTLINED ON PAGE 14, SECTION D. SAFETY CHECKLIST: RECOMMENDED SAFETY AND FUNCTION CHECKS FOR THE FRAME, CAMBER TUBES AND CROSSBRACE TO BE PERFORMED EVERY 6 MONTHS. IT ALSO STATES THAT THE USER OR CAREGIVER SHOULD PERFORM THESE WEEKLY AND MONTHLY CHECKS TO MAINTAIN THE SAFETY OF THEIR CHAIR. IF AN ITEM IS NOT WORKING PROPERLY, PLEASE CONTACT YOUR AUTHORIZED DEALER. DISCUSSION: CROSS-FRAME BREAKAGE DURING TRANSFER HAS THE POTENTIAL TO LEAD TO A FALL WHICH IS CLASSIFIED AS A POTENTIAL FOR SERIOUS INJURY OR DEATH. THEREFORE, THIS IS A REPORTABLE EVENT. CONCLUSION: ALTHOUGH THE END USER DID NOT SUSTAIN INJURIES THAT WERE LIFE THREATENING OR REQUIRED MEDICAL INTERVENTION, THERE IS POTENTIAL FOR SERIOUS INJURY WERE THIS INCIDENT TO RECUR. THEREFORE, OUT OF AN ABUNDANCE OF CAUTION, SUNRISE MEDICAL HAS DECIDED TO FILE AN MDR TO REPORT THIS INCIDENT. ADDITIONAL INFORMATION: THIS COMPLAINT HAS BEEN RE-REVIEWED AS A PART OF A FOUR-YEAR RETROSPECTIVE REVIEW AND REMEDIATION EFFORT BASED ON ENHANCEMENTS MADE TO THE COMPANY'S COMPLAINT HANDLING AND ADVERSE EVENT REPORTING PROCESSES. A LEGAL CONSULTING FIRM WAS CONSULTED FOR THE RETROSPECTIVE REVIEW. THIS MDR IS BEING FILED BASED ON THE OUTCOME OF THAT RETROSPECTIVE REVIEW.

Description of Event or Problem · 1

PER DEALER KEN, X-TUBE BROKE AT THE WELD ON THE LEFT SIDE UNDER THE SEAT. THE USER FELL WHILE TRYING TO TRANSFER FROM A CHAIR TO HIS WHEELCHAIR. THE END USER SUFFERED MINOR CUTS TO HIS LEFT LEG WITH SOME MINOR BLEEDING. THERE WAS SOME BRUISING AND PAIN TO HIS BUTTOCKS AT THE TIME OF THE FALL. HE DID NOT GO TO THE HOSPITAL TO SEEK MEDICAL ATTENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1126134 QUICKIE LXI MANUAL WHEELCHAIR IOR SUNRISE MEDICAL PRIVADA MISIONES EILXI

Patients

Seq Age Sex Outcome Treatment
1 60 YR Other