FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 1222906 · Received November 6, 2008

Report

Report Number
2023826-2008-01370
Event Type
Malfunction
Date Received
November 6, 2008
Date of Event
October 14, 2008
Report Date
October 16, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL CODES: RESULTS - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC AND ONE HAPTIC TORN. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. IT SHOULD BE NOTED THAT THE INJECTOR, CARTRIDGE AND FOAM TIP PLUNGER WERE NOT RETURNED FOR EVAL. CONCLUSIONS: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 05. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CANNOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MFG OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON ATTEMPTED TO INSERT A CC4204BF COLLAMER SINGLE PIECE LENS, BUT THE LENS WAS TORN ADVANCING THROUGH THE INJECTOR. THERE WAS NO PT CONTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL CO. CC4204BF NA

Patients

Seq Age Sex Outcome Treatment
1 INJECTOR MODEL INDIGO-P - LOT NUMBER UNK| CARTRIDGE MODEL SFC-25 FP - LOT NUMBER UNK| FOAM TIP PLUNGER MODEL FTP INDIGO - LOT NUMBER UNK