COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL
Report
- Report Number
- 2023826-2008-01370
- Event Type
- Malfunction
- Date Received
- November 6, 2008
- Date of Event
- October 14, 2008
- Report Date
- October 16, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P990013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
EVAL CODES: RESULTS - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC AND ONE HAPTIC TORN. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. IT SHOULD BE NOTED THAT THE INJECTOR, CARTRIDGE AND FOAM TIP PLUNGER WERE NOT RETURNED FOR EVAL. CONCLUSIONS: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 05. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CANNOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MFG OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.
THE REPORTER STATED THE SURGEON ATTEMPTED TO INSERT A CC4204BF COLLAMER SINGLE PIECE LENS, BUT THE LENS WAS TORN ADVANCING THROUGH THE INJECTOR. THERE WAS NO PT CONTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | CC4204BF | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INJECTOR MODEL INDIGO-P - LOT NUMBER UNK| CARTRIDGE MODEL SFC-25 FP - LOT NUMBER UNK| FOAM TIP PLUNGER MODEL FTP INDIGO - LOT NUMBER UNK |