FDA Adverse Event
Injury
Summary report: N
PINN MAR +4 10D 36IDX52OD
MDR report key: 1222449
·
Received November 4, 2008
Report
- Report Number
- 1818910-2008-05100
- Event Type
- Injury
- Date Received
- November 4, 2008
- Date of Event
- October 21, 2008
- Report Date
- October 21, 2008
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- LPH
- PMA / PMN Number
- K033273
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
EXAMINATION WAS NOT POSSIBLE, AS THE DEVICES WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT CODES SINCE THEIR RELEASE FOR DISTRIBUTION. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT ERROR WITH REGARD TO THE REPORTED EVENT. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED.
Description of Event or Problem · 1
PT WAS REVISED TO ADDRESS DISASSOCIATION OF THE LINER FROM THE CUP RESULTING IN METALOSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PINN MAR +4 10D 36IDX52OD | 87LPH | LPH | DEPUY ORTHOPAEDICS, INC. | NA | Z37B71000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Required Intervention |