FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM HYDRO

MDR report key: 1222408 · Received November 5, 2008

Report

Report Number
2953200-2008-01009
Event Type
Injury
Date Received
November 5, 2008
Date of Event
October 6, 2008
Report Date
October 6, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P990020
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: RESULTS AND CONCLUSION: (ARTERIAL TRAUMA/DISSECTION/PERFORATION); (SEVERELY CALCIFIED AND FRAIL VESSELS); (BALLOON). SECONDARY INTERVENTION REQUIRED.

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS INSERTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF ABDOMINAL AORTIC ANEURYSM. ANEURYSM MORPHOLOGY WAS NOT REPORTED. VESSEL MORPHOLOGY WAS REPORTED AS SEVERELY CALCIFIED AND FRAIL VESSELS. IT WAS REPORTED THAT PHYSICIAN WAS MODELING THE STENT GRAFT (MDR2953200-2008-01008) WITH A RELIANT BALLOON. WHILE INFLATING THE BALLOON THE VESSEL WALL WAS PERFORATED. THE BALLOON WAS COMPLETELY WITHIN THE STENT GRAFT WHILE THE BALLOON WAS BEING INFLATED. THE PHYSICIAN PLACED AN AORTIC CUFF TO TREAT THE PERFORATED VESSEL. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM HYDRO MIH MEDTRONIC CARDIOVASCULAR NA V00141150

Patients

Seq Age Sex Outcome Treatment
1 81 YR Required Intervention