FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM

MDR report key: 12222598 · Received July 24, 2021

Report

Report Number
0002023141-2021-01911
Event Type
Injury
Date Received
July 24, 2021
Date of Event
April 27, 2021
Report Date
November 13, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024020085
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER: (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. B5: DESCRIBE EVENT OR PROBLEM. D9: DEVICE AVAILABILITY. G3: DATE RECEIVED BY MANUFACTURER. G6: TYPE OF REPORT. H1: TYPE OF REPORTABLE EVENT. H2: FOLLOW UP TYPE. H6: ADVERSE EVENT PROBLEM. H10: ADDITIONAL NARRATIVE. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER: (1221084). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD (OP150) WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER: (1221084) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET (B)(4). FAX NUMBER: NOT PROVIDED. PMA/510(K) NUMBER: ADDITIONAL NUMBERS: K011028 K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THE IMPLANT AT TOOTH SITE #13 WAS REMOVED DUE TO PERI-IMPLANTITIS, INFLAMMATION AND ABSCESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1119716 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVWB8 1221084 00889024020085

Patients

Seq Age Sex Outcome Treatment
1 63 YR Male Required Intervention