FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 12222468
·
Received July 24, 2021
Report
- Report Number
- 9610877-2021-10389
- Event Type
- Malfunction
- Date Received
- July 24, 2021
- Date of Event
- March 22, 2021
- Report Date
- July 24, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- ODG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY : IT WAS CAUSED DUE TO THE RUBBER SEAL AND THE O-RING ON THE SUCTION VALVE WORN OUT. THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL OF-B65 IS AVAILABLE IN THE USA WITH A 510K NUMBER K023401. THIS REPORT IS BEING FILED AS PART OF THE PENTAX BACKLOG MANAGEMENT PLAN.
Description of Event or Problem · 1
VALVE SOMETIMES STAY FIXED IN SPITE OF THE USAGE OF SILICON OIL. THIS EVENT OCCURRED AT THE TIME OF DURING USE. THERE WAS NO REPORT OF PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1119445 | PENTAX | SUCTION/ SUCTION CHANNEL/ CONTROL VALVE FOR EG34-J10U | ODG | HOYA CORPORATION PENTAX TOKYO OFFICE | OF-B206 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |