ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2021-02049
- Event Type
- Malfunction
- Date Received
- July 23, 2021
- Report Date
- December 7, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- UDI-DI
- 10811877011320
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION WITH CORRECT LOT 1013549. THE INVESTIGATION WAS PREVIOUSLY SUBMITTED WITH LOT "1013547" WHICH AN DATA ENTRY ERROR. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 191-000 / LOT 1013549 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-430 / LOT 1013549. TEST BASE PART NUMBER 191-000 / LOT 1013549. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1013549 SHOWED THAT THE COMPLAINT RATE IS 0.04% . IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION. REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND HAVE NOT IDENTIFIED A PRODUCT DEFICIENCY. BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED CLOSED. THE PRODUCT WILL CONTINUE TO BE MONITORED AND TRACKED.
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 191-000 / LOT 1013547 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-430 / LOT 1013547. TEST BASE PART NUMBER 191-000 / LOT 1013547. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1022166 SHOWED THAT THE COMPLAINT RATE IS (B)(4). . IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION. REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND HAVE NOT IDENTIFIED A PRODUCT DEFICIENCY. BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED CLOSED. THE PRODUCT WILL CONTINUE TO BE MONITORED AND TRACKED.
REFERNCE NUMBER : 1221359-2021-02105, 1221359-2021-02106, 1221359-2021-02107, 1221359-2021-02108. THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.
THE CUSTOMER REPORTED FIVE (5) FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY ( DIFFERENT LOT NUMBERS). THIS MFR ADDRESSES PATIENT ONE (1) OF FIVE (5). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON UNKNOWN SAMPLE TYPE. REPEAT TESTING WAS NOT PERFORMED. PCR CONFIRMATION TESTING USING ALINITY ON A PHYSIOLOGIC + VIROCULT VIRAL TRANSPORT MEDIA GENERATED NEGATIVE RESULTS. THE CUSTOMER STATED THE PATIENT WAS SYMPTOMATIC EXPERIENCING ABDOMINAL PAIN THE PATIENT WAS SUPPOSED TO HAVE A FIBROSCOPY AND A COLONSCOPY. THE CUSTOMER STATED THE PATIENT WAS PLACE IN ISOLATION FOR SEVEN (7) DAYS AT HOME DUE TO THE POSITIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1118867 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 1013549 | 10811877011320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Male |