NXSTAGE SYSTEM ONE

MDR report key: 1222117 · Received November 5, 2008

Report

Report Number
3003464075-2008-00552
Event Type
Injury
Date Received
November 5, 2008
Date of Event
October 13, 2008
Report Date
October 14, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK OF THE PT'S BLOOD AS INSTRUCTED IN THE USER'S GUIDE. THE EXACT CAUSE OF THE ALARM IS NOT KNOWN. THE USER'S GUIDE INCLUDES PROBABLE CAUSES AND TROUBLESHOOTING STEPS FOR ALARMS. OCCASIONAL ALARMS DURING DIALYSIS ARE EXPECTED AND SHOULD NOT RESULT IN A BLOOD LOSS IF DEVICE LABELING IS FOLLOWED. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. NO SIMILAR PROBLEMS REPORTED SUBSEQUENT TO THIS EVENT. FACILITY STAFF HAS REVIEWED RINSEBACK PROCEDURES WITH THE OPERATOR. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A SYSTEM ALARM OCCURRED TOWARD THE END OF A ROUTINE HEMODIALYSIS TREATMENT. RINSEBACK WAS NOT PERFORMED, DUE TO THE OPERATOR NOT FOLLOWING THE PROPER RINSEBACK INSTRUCTIONS, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8087702

Patients

Seq Age Sex Outcome Treatment
1 69 YR Other