FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1222113 · Received November 5, 2008

Report

Report Number
3003464075-2008-00556
Event Type
Injury
Date Received
November 5, 2008
Date of Event
October 8, 2008
Report Date
October 8, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK IN A TIMELY MANNER FOLLOWING AN UNRECOVERABLE ALARM. THE DISPOSABLE CARTRIDGE WAS NOT AVAILABLE FOR EVALUATION. THE CYCLER WILL BE EVALUATED UPON RETURN. THE CAUSE OF THE ALARM IS NOT KNOWN AT THE TIME OF THIS REPORT. THE USER'S GUIDE CONTAINS ADEQUATE INFO REGARDING PROBABLE CAUSES OF ALARMS AND ALARM RECOVERY INSTRUCTIONS. OCCASIONAL ALARMS DURING DIALYSIS ARE EXPECTED AND SHOULD NOT RESULT IN A BLOOD LOSS IF DEVICE LABELING IS FOLLOWED. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A SYSTEM ALARM OCCURED DURING A ROUTINE HEMODIALYSIS TREATMENT WHICH COULD NOT BE RESOLVED BY THE OPERATOR. RINSEBACK WAS NOT PERFORMED DUE TO CLOTTING OF THE CIRCUIT, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 UNK

Patients

Seq Age Sex Outcome Treatment
1 68 YR Other