DIGNICAP DELTA SYSTEM
Report
- Report Number
- 3012146457-2021-00001
- Event Type
- Injury
- Date Received
- July 23, 2021
- Date of Event
- May 11, 2021
- Report Date
- July 23, 2021
- Manufacturer
- DIGNITANA INC.
- Product Code
- PMC
- PMA / PMN Number
- K191166
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NURSE
Narratives
INFORMAL TESTING OF ACTUAL UNIT 21-MAY-2021: ON 5/21/2021, DIGNITANA CLINICAL ACCOUNT MANAGER, (B)(6), WENT TO INCIDENT SITE ((B)(6) HOSPITAL) AND INSPECTED THE ACTUAL DEVICE (SN (B)(4)). NO VISIBLE FLUID WAS OBSERVED UPON ARRIVAL. MACHINE WAS MOVED FROM THE HALLWAY TO THE INSIDE OF TREATMENT ROOM (BAY 7) AND CONDUCTED A HOSE DROP TEST AND FLUID SPILLED FROM THE END OF THE THERAPY HOSE OF WHAT APPEARED TO BE CONDENSATION. FORMAL TESTING OF SIMILAR DEVICE - SAME MODEL 28-MAY-2021: ENGINEERING TESTING CONDUCTED ON A SIMILAR MODEL (DIGNICAP DELTA) BUT NOT THE ACTUAL DEVICE, FOR ONE FULL TREATMENT CYCLE CONCLUDED THAT WATER/CONDENSATION BUILD UP ON THE CONNECTORS (CONNECTING THE THERAPY HOSE TO THE DEVICE) AND INSULATION MATERIAL ON THE THERAPY HOSE COULD BE PRESENT IN QUANTITIES THAT WOULD BE ENOUGH TO WET A SURFACE NEAR THE MACHINE. NO LEAKS WERE FOUND INSIDE THE UNIT, NOR WERE ANY COOLANT LEAKS FOUND. IT IS LIKELY THAT ENVIRONMENTAL FACTORS, TEMPERATURE AND HUMIDITY CONTRIBUTED TO THE BUILD-UP OF CONDENSATION.
DIGNITANA, INC. WAS MADE AWARE ON 5/18/2021 BY STAFF AT (B)(6) CANCER CENTER ((B)(6)) THAT A NURSE SLIPPED AND FRACTURED HER ANKLE. THE INCIDENT OCCURRED ON (B)(6) 2021. A SCALP COOLING TREATMENT TO HELP REDUCE ALOPECIA IN CANCER PATIENTS, DELIVERED BY THE DIGNICAP® DELTA SYSTEM, HAD BEEN ADMINISTERED TO A PATIENT. AFTER THE TREATMENT THE DELTA CONTROL UNIT (CU) WAS MOVED OUTSIDE THE TREATMENT ROOM TO A HALLWAY. IT IS NOT KNOWN IF THE UNIT WAS WIPED DOWN AFTER THE TREATMENT. A NURSE THEN WALKED AROUND THE DEVICE, SLIPPED AND FRACTURED HER FOOT. IN A FOLLOW-UP ON (B)(6) 2021, THE NURSE STATED THAT HER ANKLE IS HEALING AND SHE RETURNED TO WORK ON (B)(6) 2021.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1116141 | DIGNICAP DELTA SYSTEM | SCALP COOLING SYSTEM | PMC | DIGNITANA INC. | DIGNICAP DELTA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |