FDA Adverse Event Injury Summary report: Y

DIGNICAP DELTA SYSTEM

MDR report key: 12220853 · Received July 23, 2021

Report

Report Number
3012146457-2021-00001
Event Type
Injury
Date Received
July 23, 2021
Date of Event
May 11, 2021
Report Date
July 23, 2021
Manufacturer
DIGNITANA INC.
Product Code
PMC
PMA / PMN Number
K191166
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INFORMAL TESTING OF ACTUAL UNIT 21-MAY-2021: ON 5/21/2021, DIGNITANA CLINICAL ACCOUNT MANAGER, (B)(6), WENT TO INCIDENT SITE ((B)(6) HOSPITAL) AND INSPECTED THE ACTUAL DEVICE (SN (B)(4)). NO VISIBLE FLUID WAS OBSERVED UPON ARRIVAL. MACHINE WAS MOVED FROM THE HALLWAY TO THE INSIDE OF TREATMENT ROOM (BAY 7) AND CONDUCTED A HOSE DROP TEST AND FLUID SPILLED FROM THE END OF THE THERAPY HOSE OF WHAT APPEARED TO BE CONDENSATION. FORMAL TESTING OF SIMILAR DEVICE - SAME MODEL 28-MAY-2021: ENGINEERING TESTING CONDUCTED ON A SIMILAR MODEL (DIGNICAP DELTA) BUT NOT THE ACTUAL DEVICE, FOR ONE FULL TREATMENT CYCLE CONCLUDED THAT WATER/CONDENSATION BUILD UP ON THE CONNECTORS (CONNECTING THE THERAPY HOSE TO THE DEVICE) AND INSULATION MATERIAL ON THE THERAPY HOSE COULD BE PRESENT IN QUANTITIES THAT WOULD BE ENOUGH TO WET A SURFACE NEAR THE MACHINE. NO LEAKS WERE FOUND INSIDE THE UNIT, NOR WERE ANY COOLANT LEAKS FOUND. IT IS LIKELY THAT ENVIRONMENTAL FACTORS, TEMPERATURE AND HUMIDITY CONTRIBUTED TO THE BUILD-UP OF CONDENSATION.

Description of Event or Problem · 1

DIGNITANA, INC. WAS MADE AWARE ON 5/18/2021 BY STAFF AT (B)(6) CANCER CENTER ((B)(6)) THAT A NURSE SLIPPED AND FRACTURED HER ANKLE. THE INCIDENT OCCURRED ON (B)(6) 2021. A SCALP COOLING TREATMENT TO HELP REDUCE ALOPECIA IN CANCER PATIENTS, DELIVERED BY THE DIGNICAP® DELTA SYSTEM, HAD BEEN ADMINISTERED TO A PATIENT. AFTER THE TREATMENT THE DELTA CONTROL UNIT (CU) WAS MOVED OUTSIDE THE TREATMENT ROOM TO A HALLWAY. IT IS NOT KNOWN IF THE UNIT WAS WIPED DOWN AFTER THE TREATMENT. A NURSE THEN WALKED AROUND THE DEVICE, SLIPPED AND FRACTURED HER FOOT. IN A FOLLOW-UP ON (B)(6) 2021, THE NURSE STATED THAT HER ANKLE IS HEALING AND SHE RETURNED TO WORK ON (B)(6) 2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1116141 DIGNICAP DELTA SYSTEM SCALP COOLING SYSTEM PMC DIGNITANA INC. DIGNICAP DELTA

Patients

Seq Age Sex Outcome Treatment
1 Other