FDA Adverse Event
Injury
Summary report: N
NXSTAGE SYSTEM ONE
MDR report key: 1221891
·
Received November 4, 2008
Report
- Report Number
- 3003464075-2008-00546
- Event Type
- Injury
- Date Received
- November 4, 2008
- Date of Event
- October 6, 2008
- Report Date
- October 6, 2008
- Manufacturer
- NXSTAGE MEDICAL, INC.
- Product Code
- KDI
- PMA / PMN Number
- K050525
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE PT'S NEEDLES DISLODGING WHILE ATTEMPTING TO ADJUST THEM. THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ON AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.
Description of Event or Problem · 1
THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. THE PT'S NEEDLES DISLODGED WHILE ADJUSTING THEM DURING A ROUTINE HEMODIALYSIS TREATMENT. TREATMENT WAS DISCONTINUED WITHOUT PERFORMING RINSEBACK, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NXSTAGE SYSTEM ONE | HIGH PERMEABILITY HEMODIALYSIS SYSTEM | KDI | NXSTAGE MEDICAL, INC. | CAR-170 | 8057701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Other |