FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1221891 · Received November 4, 2008

Report

Report Number
3003464075-2008-00546
Event Type
Injury
Date Received
November 4, 2008
Date of Event
October 6, 2008
Report Date
October 6, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE PT'S NEEDLES DISLODGING WHILE ATTEMPTING TO ADJUST THEM. THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ON AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. THE PT'S NEEDLES DISLODGED WHILE ADJUSTING THEM DURING A ROUTINE HEMODIALYSIS TREATMENT. TREATMENT WAS DISCONTINUED WITHOUT PERFORMING RINSEBACK, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8057701

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other