NXSTAGE SYSTEM ONE
Report
- Report Number
- 3003464075-2008-00548
- Event Type
- Injury
- Date Received
- November 4, 2008
- Date of Event
- October 7, 2008
- Report Date
- October 7, 2008
- Manufacturer
- NXSTAGE MEDICAL, INC.
- Product Code
- KDI
- PMA / PMN Number
- K050525
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THERE IS NO EVIDENCE OF A DEVICE MALFUNCTION. THE REPORTED BLOOD LOSS IS ATTRIBUTED TO USER ERROR AS THE OPERATOR INADVERTENTLY LEFT THE VENOUS LINE CONNECTED TO THE SALINE BAG WHEN ENTERING TREATMENT MODE. THE USER'S GUIDE PROVIDES ADEQUATE INSTRUCTIONS FOR MAKING PT CONNECTIONS WHEN ENTERING TREATMENT. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.
THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. AT THE BEGINNING OF A ROUTINE HEMODIALYSIS TREATMENT, THE OPERATOR INADVERTENTLY LEFT THE VENOUS LINE CONNECTED TO THE SALINE BAG, CAUSING IT TO FILL WITH BLOOD AS THE TREATMENT WAS STARTED. RINSEBACK WAS NOT PERFORMED RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NXSTAGE SYSTEM ONE | HIGH PERMEABILITY HEMODIALYSIS SYSTEM | KDI | NXSTAGE MEDICAL, INC. | CAR-170 | 8077717 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Other |