FDA Adverse Event
Malfunction
Summary report: N
MANUAL ORTHOPEDIC SURGICAL INSTRUMENT
MDR report key: 1221585
·
Received October 31, 2008
Report
- Report Number
- 1030489-2008-00608
- Event Type
- Malfunction
- Date Received
- October 31, 2008
- Date of Event
- October 3, 2008
- Report Date
- October 3, 2008
- Manufacturer
- WARSAW ORTHOPEDIC INC.
- Product Code
- HXX
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO MEDTRONIC FOR EVALUATION. VISUAL EXAMINATION CONFIRMED THE HEX PORTION OF THE INSTRUMENT WAS SHEARED IN THE AREA OF THE INNER SHAFT NEAR THE TRANSITION TO THE MAIN SHAFT. A REVIEW OF THE DEVICE HISTORY RECORDS FOUND NO DOCUMENTATION TO INDICATE A NON-CONFORMANCE TO SPECIFICATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE TIP OF THE DRIVER WAS BROKEN OFF WHILE TIGHTENING THE SCREW. THE TIP WAS LEFT FLUSH IN THE SCREW HEAD AND REMAINS IN THE PT. THERE WERE NO PT COMPLICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MANUAL ORTHOPEDIC SURGICAL INSTRUMENT | DRIVER | HXX | WARSAW ORTHOPEDIC INC. | NA | MC03M001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Other |