FDA Adverse Event Malfunction Summary report: N

30 ML BD LUER-LOK TIP SYRINGE

MDR report key: 12215078 · Received July 22, 2021

Report

Report Number
1911916-2021-00722
Event Type
Malfunction
Date Received
July 22, 2021
Date of Event
June 24, 2021
Report Date
July 8, 2021
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903028321
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED AN ANT WAS FOUND IN THE BD 30ML SYRINGE AFTER MIXING AND ASPIRATING SOLUTION FROM A VIAL. TO AID IN THE INVESTIGATION, TWO PHOTOS WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. BOTH PHOTOS SHOW A SYRINGE CONNECTED TO A NEEDLE ASSEMBLY. THE SYRINGE HAS SOLUTION IN IT AND A FOREIGN MATTER. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302832, LOT NUMBER 0062957. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED, BUT WITH NO PHYSICAL SAMPLE ANALYSIS A PROBABLE ROOT CAUSE COULD NOT BE OFFERED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING AND ASPIRATING SOLUTION THE 30 ML BD LUER-LOK¿ TIP SYRINGE WAS FOUND TO HAVE FOREIGN MATTER IN THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: AN ANT (INSECT) WAS FOUND IN THE BD 30ML SYRINGE AFTER MIXING AND ASPIRATING SOLUTION FROM CLOXACILLIN VIAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1105426 30 ML BD LUER-LOK TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 302832 0062957 30382903028321

Patients

Seq Age Sex Outcome Treatment
1