COBAS C111
Report
- Report Number
- 1823260-2008-08185
- Event Type
- Malfunction
- Date Received
- November 5, 2008
- Date of Event
- October 15, 2008
- Report Date
- November 5, 2008
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K071211
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 129, REPEAT 142. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 129, REPEAT 140. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 131, REPEAT 142. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 127, REPEAT 139. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 132, REPEAT 143. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL SODIUM RESULT 129, REPEAT 142; INITIAL POTASSIUM RESULT 4.0, REPEAT 4.7. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 128, REPEAT 136. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 130, REPEAT 140. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 127, REPEAT 136. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 130, REPEAT 139. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
USER RECEIVED LOW SODIUM AND POTASSIUM RESULTS FOR 11 PT SAMPLES. ALL RESULTS IN MMOL/L. REPEAT TESTING WAS PERFORMED IN 2008. SAMPLE INITIAL RESULT 129, REPEAT 139. SAMPLES WERE NOT REPORTED OUT UNTIL TROUBLESHOOTING WAS COMPLETED. USER RECALIBRATED WITH NEW ISE SOLUTIONS WHICH IMPROVED THE PERFORMANCE OF THE ASSAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS C111 | CLINICAL CHEMISTRY ANALYZER-JJE | JJE | ROCHE DIAGNOSTICS |