CELL-DYN 1700 CS ANALYZER
Report
- Report Number
- 2919069-2008-00770
- Event Type
- Malfunction
- Date Received
- November 7, 2008
- Date of Event
- October 6, 2008
- Report Date
- October 16, 2008
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION/CELLDYN
- Product Code
- GKZ
- PMA / PMN Number
- K870233
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
(B)(4). A FIELD SERVICE REPRESENTATIVE (FSR) WAS DISPATCHED TO THE ACCOUNT. THE FSR CHANGED THE PERIPUMP TUBING AND CLEANED THE CLOSED MODE. THE FSR ALSO REPLACED A WORN OUT CLOSED SAMPLE DETECTOR ON THE CAP PIERCER, PERFORMED SAMPLE SENSOR ALIGNMENT, REPLACED A CLOGGED NEEDLE DRIVE ASSEMBLY, AND VERIFIED ALL GAINS. PRECISION AND QC WERE RUN AND WERE WITHIN SPECIFICATIONS. LABELING REVIEW THE CELL-DYN 1700 SYSTEM OPERATOR'S MANUAL, LIST NUMBER 03H58-01, REVISION F, UNDER SECTION 9, SERVICE AND MAINTENANCE, NONSCHEDULED MAINTENANCE FREQUENCY, PAGE 9-19, INDICATES THAT THE SAMPLE ASPIRATION PROBE INTERIOR SHOULD BE CLEANED WHEN IT IS SUSPECTED OF BEING THE SOURCE OF IMPRECISION. ASPIRATION PROBE REMOVAL AND REPLACEMENT SHOULD BE PERFORMED WHEN THERE IS A NON-REMOVABLE OBSTRUCTION IN THE PROBE. SECTION 10, TROUBLESHOOTING AND DIAGNOSTICS, PAGE 10-9, PROVIDES BASIC TROUBLESHOOTING TO ASSIST THE OPERATOR IN IDENTIFYING, ISOLATING AND RESOLVING INSTRUMENT PROBLEMS. SECTION 13, CELL-DYN 1700CS, CLOSED SAMPLE ASPIRATION, SYSTEM COMPONENTS, CLOSED SAMPLE ASSEMBLY, PAGES 13-3 THROUGH 13-4, PROVIDES INFORMATION IN REGARDS TO THE CLOSED SAMPLE ASPIRATION ASSEMBLY AND ITS EXTERIOR AND INTERIOR COMPONENTS. A REVIEW OF THE COMPLAINT ANALYSIS TRENDING SYSTEM (CATS) AND TRENDING ANALYSIS SUPPORT SYSTEM (TASS) REPORTS WAS PERFORMED AND NO ADVERSE TREND RELATED TO DISCREPANT RESULTS WITH PATIENT SAMPLES IN CLOSED MODE WAS IDENTIFIED. NO MALFUNCTION WAS IDENTIFIED. THIS IS THE FINAL REPORT.
THE ACCOUNT STATED THAT THE CELL-DYN 1700 ANALYZER GENERATED THE FOLOWING RESULTS IN THE CLOSED MODE: WBC=3,400/UL, RBC=2.85X10E6/UL, HGB=8.4 G/DL, PLT=145,000/UL. THE SAMPLE WAS REPEATED IN THE OPEN MODE WITH THE FOLLOWING RESULTS: WBC=4,000/UL, RBC=3.37X10E6/UL, HGB=9.9 G/DL, PLT=180,000/UL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL-DYN 1700 CS ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION/CELLDYN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |