FDA Adverse Event Malfunction Summary report: N

DEPTH GAUGE FOR 3.5MM CORTEX SCREWS

MDR report key: 12214159 · Received July 22, 2021

Report

Report Number
2939274-2021-04231
Event Type
Malfunction
Date Received
July 22, 2021
Report Date
June 24, 2021
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTJ
UDI-DI
10886982190000
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6- THIS COMPLAINT WAS NOT CONFIRMED DURING THE INVESTIGATION AS THE RETURNED DEVICE WAS NOT BROKEN IN TWO PIECES. HOWEVER, THE TIP OF THE NEEDLE COMPONENT WAS OBSERVED TO BE BENT. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED BASED ON THE PROVIDED INFORMATION. NO NEW, UNIQUE, OR DIFFERENT PATIENT HARMS WERE IDENTIFIED AS A RESULT OF THIS EVALUATION. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. NO DESIGN ISSUES WERE OBSERVED DURING THE DOCUMENT/SPECIFICATION REVIEW. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT : PART # 319.091, SYNTHES LOT # 6191217, SUPPLIER LOT # N/A, RELEASE TO WAREHOUSE DATE: 29 JUL 2009, MANUFACTURING LOCATION: BRANDYWINE, NO NCR'S WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

OCCUPATION: REPORTER IS A SYNTHES EMPLOYEE. THE DEVICE WAS RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO A CASE ON AN UNKNOWN DATE, THE DEPTH GAUGE TIP WAS BROKEN OFF WHILE BEING ASSEMBLED. THIS REPORT IS FOR A DEPTH GAUGE FOR 3.5MM CORTEX SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1110116 DEPTH GAUGE FOR 3.5MM CORTEX SCREWS GAUGE, DEPTH HTJ WRIGHTS LANE SYNTHES USA PRODUCTS LLC 319.091 6191217 10886982190000

Patients

Seq Age Sex Outcome Treatment
1