FDA Adverse Event Malfunction Summary report: N

IMAGIQ

MDR report key: 1220943 · Received November 4, 2008

Report

Report Number
1220943
Event Type
Malfunction
Date Received
November 4, 2008
Date of Event
November 3, 2008
Report Date
November 4, 2008
Manufacturer
STILLE SURGICAL
Product Code
KXJ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TN, US

Narratives

Description of Event or Problem · 1

DOCTOR FINISHING A LEFT FEMORAL POPLITEAL BYPASS, STAPLING LEG, WHEN THE FOOT OF THE TABLE STARTED DRIFTING DOWN. THE TABLE WAS LEVELED AND A STOOL WAS PLACED UNDER THE LEG TO FINISH THE PROCEDURE. PATIENT WAS UNHARMED AND UNIT WAS SENT TO BIOMEDICAL SERVICES. EXAMINATION OF TABLE FOUND THAT A HYDRAULIC HOSE FEEDING THE TREND/REVERSE TREND CYLINDER HAS BEEN DAMAGED FROM RUBBING AGAINST THE SUPPORT BLADE USED TO SUPPORT THE HOSES FEEDING THE UPPER PORTION OF THE TABLE. THE TABLE IS BEING HELD FOR INSPECTION BY MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMAGIQ IMAGING TABLE KXJ STILLE SURGICAL IMAGIQ *

Patients

Seq Age Sex Outcome Treatment
1 75 YR