FDA Adverse Event
Malfunction
Summary report: N
IMAGIQ
MDR report key: 1220943
·
Received November 4, 2008
Report
- Report Number
- 1220943
- Event Type
- Malfunction
- Date Received
- November 4, 2008
- Date of Event
- November 3, 2008
- Report Date
- November 4, 2008
- Manufacturer
- STILLE SURGICAL
- Product Code
- KXJ
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- TN, US
Narratives
Description of Event or Problem · 1
DOCTOR FINISHING A LEFT FEMORAL POPLITEAL BYPASS, STAPLING LEG, WHEN THE FOOT OF THE TABLE STARTED DRIFTING DOWN. THE TABLE WAS LEVELED AND A STOOL WAS PLACED UNDER THE LEG TO FINISH THE PROCEDURE. PATIENT WAS UNHARMED AND UNIT WAS SENT TO BIOMEDICAL SERVICES. EXAMINATION OF TABLE FOUND THAT A HYDRAULIC HOSE FEEDING THE TREND/REVERSE TREND CYLINDER HAS BEEN DAMAGED FROM RUBBING AGAINST THE SUPPORT BLADE USED TO SUPPORT THE HOSES FEEDING THE UPPER PORTION OF THE TABLE. THE TABLE IS BEING HELD FOR INSPECTION BY MANUFACTURER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMAGIQ | IMAGING TABLE | KXJ | STILLE SURGICAL | IMAGIQ | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR |