LIFEVEST WCD 3000 SYSTEM
Report
- Report Number
- 3002158293-2008-00569
- Event Type
- Malfunction
- Date Received
- October 29, 2008
- Date of Event
- October 14, 2008
- Report Date
- October 29, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF THE ELECTRODE BELT HAS BEEN COMPLETED. UPON FURTHER INSPECTION, THE ELECTRODE BELT HAD PINS THAT WERE BENT INSIDE THE CONNECTOR. THERE WERE NO PINS MISSING. THE ROOT CAUSE OF THE BENT PINS WAS PROBABLY EXCESSIVE FORCE APPLIED TO THE CONNECTOR DURING MATING. THE CONNECTOR WAS FORCED INTO THE MONITOR WHICH DEFEATED THE CONNECTOR KEYING MECHANISM AND ALLOWED THE PINS TO MISALIGN WITH THE MATING SOCKETS. THE DAMAGED ELECTRODE BELT CONNECTOR WAS REPLACED. IT WAS RETESTED AND THEN RESTOCKED. NO ADVERSE EVENT RESULTED FROM THE BENT PINS. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT EXPERIENCE ANY PROBLEMS WITH IT.
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING OF AN ELECTRODE BELT, WHICH WAS RETURNED FOR ROUTINE MAINTENANCE, IT WAS DISCOVERED THAT THE CONNECTOR PINS WERE BENT. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT EXPERIENCE ANY PROBLEMS WITH IT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |