FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 1220880 · Received October 29, 2008

Report

Report Number
3002158293-2008-00569
Event Type
Malfunction
Date Received
October 29, 2008
Date of Event
October 14, 2008
Report Date
October 29, 2008
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF THE ELECTRODE BELT HAS BEEN COMPLETED. UPON FURTHER INSPECTION, THE ELECTRODE BELT HAD PINS THAT WERE BENT INSIDE THE CONNECTOR. THERE WERE NO PINS MISSING. THE ROOT CAUSE OF THE BENT PINS WAS PROBABLY EXCESSIVE FORCE APPLIED TO THE CONNECTOR DURING MATING. THE CONNECTOR WAS FORCED INTO THE MONITOR WHICH DEFEATED THE CONNECTOR KEYING MECHANISM AND ALLOWED THE PINS TO MISALIGN WITH THE MATING SOCKETS. THE DAMAGED ELECTRODE BELT CONNECTOR WAS REPLACED. IT WAS RETESTED AND THEN RESTOCKED. NO ADVERSE EVENT RESULTED FROM THE BENT PINS. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT EXPERIENCE ANY PROBLEMS WITH IT.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING OF AN ELECTRODE BELT, WHICH WAS RETURNED FOR ROUTINE MAINTENANCE, IT WAS DISCOVERED THAT THE CONNECTOR PINS WERE BENT. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT EXPERIENCE ANY PROBLEMS WITH IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3100 NA

Patients

Seq Age Sex Outcome Treatment
1 NA