OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2008-00273
- Event Type
- Malfunction
- Date Received
- October 29, 2008
- Date of Event
- October 2, 2008
- Report Date
- October 2, 2008
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PRODUCT WILL NOT BE RETURNED SO NO EVALUATION IS POSSIBLE. THE CUSTOMER NOTICED A "CRACKING" NOISE DURING THE FILL PROCESS INDICATES A PROBABLE PROBLEM WITH THE RETAINER THAT ALLOWED A COMPONENT TOO MOVE RESULTING IN DAMAGE. THIS DAMAGE PREVENTED THE PLUNGER FROM ADVANCING. THE USER WOULD HAVE BEEN AWARE OF A PROBLEM DURING THE FILL PROCESS. THERE IS RESISTANCE FELT IN FILLING THE POD WITH INSULIN AND A DISTINCT "CRACKLING" NOISE RESULTING FROM STRIPPING THE THREADS OF THE COMPONENT WHEN OVER-PRESSURIZED. THE OMNIPOD USER GUIDE STATES: "WARNING: NEVER USE A POD IF, DURING FILL, YOU DETECT ANY CRACKLING NOISE OR RESISTANCE WHILE DEPRESSING THE PLUNGER OF THE FILL SYRINGE. USING A POD WITH THESE CONDITIONS COULD RESULT IN UNDER-DELIVERY OF INSULIN." THE USER IS ALSO INSTRUCTED IN THE USER GUIDE TO MONITOR BLOOD GLUCOSE LEVELS FREQUENTLY. BY FOLLOWING THESE RECOMMENDATIONS, THE USER BECAME AWARE OF THEIR HIGH BLOOD GLUCOSE AND STARTED A NEW POD OR BACKUP THERAPY IF NEEDED.
CUSTOMER CALLED TO REPORT A POD THAT WAS HARD TO FILL AND HE HEARD CRACKLING NOISES DURING FILL. CUSTOMER ACTIVATED THE POD ANYWAY AND DURING POD WEAR, CUSTOMER RECEIVED BLOOD GLUCOSE (BG) READINGS OF "HI" AND HAD SEVERE HYPERGLYCEMIC SYMPTOMS EVEN AFTER SEVERAL INSULIN BOLUSES. REVIEWED LABELLING INSTRUCTIONS WITH THE CUSTOMER. NO FURTHER ISSUES WERE IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L11644 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Other |