670G INSULIN PUMP MMT-1780KL
Report
- Report Number
- 2032227-2021-170738
- Event Type
- Malfunction
- Date Received
- July 22, 2021
- Date of Event
- July 12, 2021
- Report Date
- October 19, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000365882
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
RETAINER RING = BLACK. DEVICE RECEIVED WITH BLANK DISPLAY DUE TO MOISTURE DAMAGE ON ELECTRONIC ASSEMBLY (PCBA 1). A WORKING TEST PCBA 1 WAS SWAPPED OUT AND THE PUMP POWERED UP AND ALARMED CRITICAL PUMP ERROR (OPEN BOOK IMAGE). CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM DUE TO MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY (PCBA 2). UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO DOWNLOAD FIRMWARE USING CREST. UNABLE TO DOWNLOAD HISTORY FILES AND TRACES USING THUS. FORCE SENSOR ZERO OFFSET NOT WITHIN SPECIFICATION. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE, CRACKED BELT CLIP RAIL CASE CORNER, PILLOWING KEYPAD OVERLAY AND CORRODED BATTERY TUBE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. CUSTOMER STATED THAT THE INSULIN PUMP GOT WET IN THE BEACH AND VIBRATED THEN THEY TOOK THE BATTERY OUT AFTER THAT THE INSULIN PUMP WAS NOT RESPONDING. THE DISPLAY WAS NOT BLANK LESS THAN 30 SECONDS AND DID NOT RETURN. CUSTOMER REPORTED THAT DISPLAY WAS BLANK CURRENTLY. INSERT BATTERY ALARM DID NOT DISPLAY ON THE INSULIN PUMP SCREEN. CUSTOMER STATED THAT THE DISPLAY DID NOT RETURN AFTER INSULIN PUMP RESTART. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1111031 | 670G INSULIN PUMP MMT-1780KL | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KL | HG4NS82 | 000000763000365882 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |