FDA Adverse Event Malfunction Summary report: N

PEN NDL 32G 4MM PRO

MDR report key: 12208070 · Received July 21, 2021

Report

Report Number
9616656-2021-00875
Event Type
Malfunction
Date Received
July 21, 2021
Date of Event
June 25, 2021
Report Date
June 30, 2021
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
K162516
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY : EXEC SUMMARY - NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AND THE ROOT CAUSE IS UNDETERMINED. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED, THIS IS THE 3RD RELATED COMPLAINT FOR NEEDLE CLOG ON LOT #: 0231157. NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. CAPA/SA - BASED ON THE ABOVE NO ADDITIONAL INVESTIGATION AND NO CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) ARE REQUIRED AT THIS TIME. DHR REVIEW - A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT 1 PEN NDL 32G 4MM HP NEEDLE WAS UNABLE TO DELIVER INSULIN OR MEDICINE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CONSUMER THE INSULIN WOULD NOT COME THROUGH THE NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1103722 PEN NDL 32G 4MM PRO HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 0231157

Patients

Seq Age Sex Outcome Treatment
1