FDA Adverse Event Injury Summary report: N

WHITESTAR SIGNATURE

MDR report key: 12207751 · Received July 21, 2021

Report

Report Number
2020664-2021-07064
Event Type
Injury
Date Received
July 21, 2021
Date of Event
June 8, 2021
Report Date
July 21, 2021
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
HQC
PMA / PMN Number
K160236
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ETHNICITY: UNKNOWN/ NOT PROVIDED. SUSPECT MEDICAL DEVICE: CUSTOMER PROVIDED POSSIBLE SUSPECT OPO73 TUBING LOTS INCLUDE THE FOLLOWING LOTS AVAILABLE IN THEIR INVENTORY - 60289899, 60279912, 60297871, 60282067; HOWEVER, CUSTOMER COULD NOT ASSOCIATE A SPECIFIC LOT TO THE CASE. THEREFORE, THE FOLLOWING FIELDS LISTED BELOW RELATED TO SECTION D AND H ARE UNKNOWN FOR THE CASE. LOT NUMBER IS UNKNOWN AS IT WAS NOT PROVIDED. EXPIRATION DATE IS UNKNOWN AS THE LOT NUMBER WAS NOT PROVIDED. UNIQUE IDENTIFIER IS UNKNOWN AS THE LOT NUMBER WAS NOT PROVIDED. MANUFACTURER DATE IS UNKNOWN AS THE LOT NUMBER WAS NOT PROVIDED. THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF THE DEVICE/LOT HISTORY RECORD AND COMPLAINT TRENDING FOR THIS DEVICE WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. THE CLINIC IS REPORTING THIS ADVERSE EVENT ONLY AND DID NOT REQUEST OR REQUIRE FIELD SERVICE OR CLINICAL SUPPORT. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE 6 CASES OF TOXIC ANTERIOR SEGMENT SYNDROME (TASS). IT WAS LEARNED THAT FOR THIS EVENT, THE VISUAL SYMPTOMS ARE NOT DEBILITATING. THE PATIENT PRESENTED WITH HISTORY OF SIGNIFICANT PAIN, BLURRED VISION SINCE THE EVENING OF SURGERY, DIFFUSE CORNEAL EDEMA, ELEVATED INTRAOCULAR PRESSURE (IOP) TO 60 MMHG, MILD ANTERIOR SEGMENT INFLAMMATION, ORIGINALLY THOUGHT EDEMA SECONDARY TO ELEVATED IOP BUT WITH LOWERING OF IOP, LIMBAL TO LIMBAL EDEMA PERSISTED AND THEN CONSIDERED TO BE TASS TREATMENT PROVIDED WAS INTENSIVE TOPICAL STEROIDS AND IOP MEDS. THE FOLLOWING MEDS WERE ALSO GIVEN BUT WAS NOTED AS THE DOCTOR'S STANDARD CARE, PROLENSA EVERYDAY, INVELTYS 2X/DAY, MOXIFLOXACIN 2X/DAY, LUMIGAN, COMBIGAN 2X/DAY, SIMBRINZA 2X/DAY, DIAMOX 500 MG TABLETS. IT WAS STATED THAT CURRENT PATIENT STATUS IS THE IOP NORMAL, CORNEAL EDEMA RESOLVED, PUPIL REMAINS ENLARGED POSSIBLY RELATED TO ELEVATED IOP WITH IRIS ISCHEMIA AND OR SECONDARY TO POSSIBLE TASS, BEST CORRECTED VISUAL ACUITY (BCVA) 20/30. INTRAOCULAR LENS (IOL) REMAINS IMPLANTED. NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTION REQUIRED. THE OTHER PRODUCTS ARE NOT BEING RETURNED FOR EVALUATION. ADDITIONAL INFORMATION RECEIVED NOTED THAT THE CUSTOMER INVESTIGATED THE CASES. HOWEVER, ROOT CAUSE FOR THE TASS HAS NOT BEEN DETERMINED. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS 6 OF 6 CASES FOR TASS CAPTURING THE EVENT FOR THE TUBING PACK. SEPARATE REPORTS ARE BEING SUBMITTED FOR EACH OF THE OTHER PRODUCTS INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1101471 WHITESTAR SIGNATURE DUAL PUMP PACK HQC JOHNSON & JOHNSON SURGICAL VISION, INC. OPO73 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention