FDA Adverse Event Injury Summary report: N

SHILEY, 8DCT 7.6 MM ID 12.2 MM OD 79.0 MM L

MDR report key: 1220684 · Received October 31, 2008

Report

Report Number
MW5008857
Event Type
Injury
Date Received
October 31, 2008
Date of Event
September 24, 2008
Report Date
October 30, 2008
Manufacturer
NELLCOR
Product Code
BTO
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
VA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

TRACHEOSTOMY BROKE OFF AT FLANGE WHILE ATTACHED TO PATIENT. PHYSICIAN WAS ABLE TO TEMPORARILY REPAIR THE DEVICE WITH A STERILE SURGICAL ADHESIVE UNTIL THE PATIENT WAS STABLE ENOUGH TO REPLACE THE TRACH. THE HUB SEPARATED FROM THE OUTER CANNULA AT THE OUTER CANNULA WAS STILL ATTACHED TO THE FLANGE BY A PIVOT ON EACH SIDE THUS PREVENTING IT FROM MIGRATING DEEPER INTO THE PATIENT'S AIRWAY, ALTHOUGH IN THIS CASE THE INNER CANNULA COULD SEPARATE FROM THE PATIENT'S AIRWAY. IN 2004, THERE WERE SIMILAR INCIDENTS WITH THE SHILEY. TRACHEOSOFT XLT EXTENDED LENGTH TRACHEOSTOMY TUBE AFTER WHICH THE PRODUCT WAS RECALLED. THE RECALL WAS PROMPTED BY REPORTS OF THE OUTER CANNULA SEPARATING FROM THE HUB AND SWIVEL NECK PLATE. IF THIS FAILURE MODE DOES OCCUR. THE OUTER CANNULA COULD TRAVEL INTO THE PATIENT'S AIRWAY, POTENTIALLY INTERFERING WITH BREATHING OR VENTILATION AND RESULTING IN SERIOUS INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SHILEY, 8DCT 7.6 MM ID 12.2 MM OD 79.0 MM L TRACHEOSTOMY TUBE BTO NELLCOR 8DCT 0802000385

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| R MAQUETTE SERVOI VENTILATOR| FISHER PAYKAL UNIVERSAL PATIENT CIRCUIT