FDA Adverse Event Injury Summary report: N

CRYSTALENS

MDR report key: 1220370 · Received November 4, 2008

Report

Report Number
2031924-2008-00316
Event Type
Injury
Date Received
November 4, 2008
Date of Event
July 7, 2008
Report Date
September 17, 2008
Manufacturer
BAUSCH & LOMB SURGICAL
Product Code
HQL
PMA / PMN Number
P030002
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORDS WERE REVIEWED AND THERE WERE NO DISCREPANCIES OR UNUSUAL FINDINGS. ACCORDING TO THE PHYSICIAN, THE ROOT CAUSE OF THE REPORTED PROBLEM OF LENS VAULTING WAS RELATED TO CAPSULAR CONTRACTION SYNDROME.

Description of Event or Problem · 1

THE PHYSICIAN REPORTS THAT A PATIENT UNDERWENT CATARACT SURGERY WITH IMPLANTATION OF THE CYSTALENS IN THE RIGHT EYE. APPROXIMATELY SIX MONTHS POSTOPERATIVELY, THE LENS VAULTED DUE TO CAPSULAR CONTRACTION SYNDROME. PREOPERATIVELY, THE PATIENT'S BCVA WAS 20/25 AND MRSE WAS +3.75 SPH. ONCE THE LENS VAULTED, THE MRSE CHANGED TO -1.00 -200 X 036. A YAG CAPSULOTOMY WAS PERFORMED TO REPOSITION THE LENS AND THE BCVA IMPROVED TO 20/25+2, J3, AND THE MRSE WAS PLANO SPH. ACCORDING TO THE PHYSICIAN, THE CAUSE OF THE EVENT WAS DUE TO CAPSULAR PHIMOSIS WHICH CAUSED THE SUPERIOR HAPTIC TO VAULT POSTERIORLY AND THE INFERIOR HAPTIC TO VAULT ANTERIORLY. THE PATIENT'S PROGNOSIS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CRYSTALENS INTRAOCULAR LENS HQL BAUSCH & LOMB SURGICAL AT50SE 007096

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention