ADULT DUAL-HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2008-00647
- Event Type
- Malfunction
- Date Received
- October 31, 2008
- Report Date
- October 3, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BTT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS IS A MALFUNCTION THAT HAS BEEN REPORTED TO FISHER & PAYKEL HEALTHCARE BY A DISTRIBUTOR. THE PRODUCT IS NOT SOLD IN THE USA, BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510 (K) FOR THAT PRODUCT IS K983112. METHOD: THE ELECTRICAL RESISTANCE OF THE HEATER WIRE IN BOTH THE INSPIRATORY AND EXPIRATORY TUBES OF THE RT200 BREATHING CIRCUIT WAS TESTED USING A MULTIMETER. RESULTS: THE INSPIRATORY HEATER WIRE WAS AN OPEN LOOP DUE TO A BREAK IN THE CONNECTION BETWEEN THE HEATER WIRE AND ONE OF THE PINS THAT CRIMPS TO THE HEATER WIRE. A LOT CHECK COULD NOT BE PERFORMED AS NO LOT NUMBER HAS BEEN PROVIDED. CONCLUSION: ELECTRICAL OPEN CIRCUITS IN HEATER WIRES ARE OFTEN ASSOCIATED WITH IMPROPER CRIMPING OF THE HEATER WIRE DURING PRODUCTION. ALL BREATHING CIRCUITS ARE ELECTRICALLY TESTED DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THE HEATER WIRE BECAME AN OPEN CIRCUIT POST PRODUCTION, POSSIBLY AS A RESULT OF A WEAK CRIMP CONNECTION. FURTHER DESIGN IMPROVEMENTS TO THE EXISTING CRIMPING MACHINES ARE IN PROGRESS. OUR MONITORING AND TRENDING OF OPEN CIRCUIT HEATER WIRES IN ADULT BREATHING CIRCUITS HAS A RATE OF OCCURRENCE WORLDWIDE.
A HEALTHCARE FACILITY REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT THE MR850 RESPIRATORY HUMIDIFIER ISSUED A HEATER WIRE ALARM WHEN CONNECTED TO AN RT200 ADULT DUAL-HEATED BREATHING CIRCUIT. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL-HEATED BREATHING CIRCUIT | BTT | FISHER & PAYKEL HEALTHCARE, LTD. | RT200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | MR850 HUMIDIFIER |