FDA Adverse Event Malfunction Summary report: N

ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 1220357 · Received October 31, 2008

Report

Report Number
9611451-2008-00648
Event Type
Malfunction
Date Received
October 31, 2008
Report Date
October 3, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BZE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE ELECTRICAL RESISTANCE OF THE HEATER WIRES IN BOTH THE INSPIRATORY AND EXPIRATORY TUBES OF THE RT340 BREATHING CIRCUITS WAS TESTED USING A MULTIMETER. RESULTS: THE INSPIRATORY HEATER WIRE IN THE CIRCUIT WAS AN OPEN LOOP DUE TO A BREAK IN THE CONNECTION BETWEEN THE HEATER WIRE AND ONE OF THE PINS THAT CRIMPS TO THE HEATER WIRE. NO LOT CHECK COULD BE PERFORMED AS NO LOT NUMBER HAS BEEN PROVIDED. CONCLUSION: ELECTRICAL OPEN CIRCUITS IN HEATER WIRES ARE OFTEN ASSOCIATED WITH IMPROPER CRIMPING OF THE HEATER WIRE DURING PRODUCTION. ALL BREATHING CIRCUITS ARE ELECTRICALLY TESTED DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THE HEATER WIRE BECAME AN OPEN CIRCUIT POST CONNECTION, POSSIBLY AS A RESULT OF A WEAK CRIMP CONNECTION. FURTHER DESIGN IMPROVEMENTS TO THE EXISTING CRIMPING MACHINES ARE IN PROGRESS. OUR MONITORING AND TRENDING OF OPEN CIRCUIT HEATER WIRES IN ADULT BREATHING CIRCUITS HAS A RATE OF OCCURRENCE IN THE LAST YEAR.

Description of Event or Problem · 1

A HOSPITAL REPORTED TO A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT THE HEATER WIRE WAS NOT FUNCTIONING IN AN RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT. REPLACING THE BREATHING CIRCUIT RESOLVED THE PROBLEM. THIS WAS DETECTED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT BZE FISHER & PAYKEL HEALTHCARE, LTD. RT340

Patients

Seq Age Sex Outcome Treatment
1