REAL INTELLIGENCE CORI
Report
- Report Number
- 3010266064-2021-00562
- Event Type
- Malfunction
- Date Received
- July 21, 2021
- Date of Event
- July 9, 2021
- Report Date
- November 3, 2022
- Manufacturer
- BLUE BELT TECHNOLOGIES
- Product Code
- OLO
- UDI-DI
- 00885556757420
- PMA / PMN Number
- K193120
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
H3, H6: THE REAL INTELLIGENCE CORI, PART NUMBER ROB10024, SERIAL NUMBER SN(B)(6), INTENDED FOR TREATMENT WAS NOT RETURNED FOR EVALUATION. AN IMAGE WAS PROVIDED. A RELATIONSHIP BETWEEN THE REPORTED EVENT AND THE DEVICE WAS ESTABLISHED. THE REPORTED PROBLEM WAS CONFIRMED WITH A VISUAL INSPECTION. THE IMAGE PROVIDED CONFIRMED THE CRITICAL ERROR DISPLAYED. A COMPLAINT HISTORY REVIEW FOR SIMILAR REPORTED/CONFIRMED COMPLAINTS HAS IDENTIFIED PRIOR EVENTS. A REVIEW OF MANUFACTURING RECORDS INDICATE THE DEVICE MET ALL SPECIFICATIONS UPON RELEASE INTO DISTRIBUTION. THE FAILURE MODE AND ASSOCIATED RISK HAVE BEEN ANTICIPATED WITHIN THE RISK FILE. THE RISK LEVEL IS STILL ADEQUATE. A HISTORICAL REVIEW CONCLUDED THAT NO PRIOR ESCALATION ACTIONS ARE APPLICABLE TO THE SCOPE OF THE REPORTED COMPLAINT. THE MOST LIKELY CAUSE OF THIS EVENT IS ASSOCIATED WITH A SOFTWARE BUG. BASED ON THE INVESTIGATION, THE NEED FOR A CORRECTIVE ACTION IS NOT RECOMMENDED OR REQUIRED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED THE COMPLAINT WILL BE REOPENED. THE FAILURE MODE WILL CONTINUE TO BE CLOSELY MONITORED THROUGH COMPLAINT INVESTIGATION AND TRENDED THROUGH POST MARKET SURVEILLANCE ACTIVITIES.
IT WAS REPORTED THAT, DURING A TEST PRIOR A CORI-LAB, A CRITICAL ERROR WAS DISPLAYED AFTER QUITTING OUT OF PASSED DRILL DIAGNOSTICS. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1101657 | REAL INTELLIGENCE CORI | ORTHOPEDIC STEREOTAXIC INSTRUMENT | OLO | BLUE BELT TECHNOLOGIES | ROB10024 | 00885556757420 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |