FDA Adverse Event Malfunction Summary report: N

CONCERTO

MDR report key: 1220264 · Received October 31, 2008

Report

Report Number
9611530-2008-00062
Event Type
Malfunction
Date Received
October 31, 2008
Date of Event
September 26, 2008
Report Date
October 30, 2008
Manufacturer
ARJO HOSPITAL EQUIPMENT AB
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN ARJO INVESTIGATOR EXAMINED THE DEVICE AND FOUND WHITE EGG SHELL PAINT MARKS ON THE SIDE RAILS. THERE WAS A CRACK IN THE PAINT ON THE RIGHT SIDE RAIL WHERE IT WAS BENT. THERE WERE SOME SMALL HOLES THROUGHOUT THE PAD, AND THE BATTERY COVER WAS CRACKED. THE SIDE RAILS WERE BENT, PREVENTING THE SIDE BRACKETS FROM LOCKING INTO PLACE. THE REST OF THE UNIT FUNCTION TESTED CORRECTLY. THE CUSTOMER STATED THEY HAD RE-BENT THE SIDE RAIL BACK INTO PLACE OVER A PERIOD OF TIME. THE EVENT HAS BEEN RECREATED; THE RIGHT SIDE RAIL WILL NOT LOCK. THE MAINTENANCE SUPERVISOR AT THE SITE ALSO STATED THAT THEY NEED HANDS-ON TRAINING FOR THEIR STAFF ON ALL UNITS. MANY OTHER SHOWER TROLLEYS WERE IMPROPERLY BEING USED BY LOADING ITEMS SUCH AS TV'S, ETC. CAUSING DAMAGE TO UNIT PADS. THE MANUFACTURER CONCLUDES THE ROOT CAUSE OF THE EVENT IS BOTH USER ERROR AND LACK OF MAINTENANCE. THE PRODUCT HAS BEEN IN NEED OF SERVICE/TECHNICAL CHECK FOR QUITE SOME TIME (FOUND IN BAD CONDITION); A DEFECT PRODUCT WAS USED. BOTH SAFETY CATCHES MUST LOCK. THE OPERATING AND PRODUCT CARE INSTRUCTIONS STATE, UNDER THE CHAPTER MAINTENANCE, TO WEEKLY EXAMINE THE SHOWER TROLLEY FOR DAMAGE AND PERFORM AN ANNUAL EXCHANGE OF THE SAFETY CATCH. THIS MOST LIKELY HAS NOT BEEN DONE, ACCORDING TO THE STATEMENTS DESCRIBING THE BAD CONDITION OF THE DEVICE. THE MANUFACTURER RECOMMENDS LIFT OPERATORS BE RETRAINED REGARDING THE VARIOUS FUNCTIONS AND USE OF THE PRODUCT. DAMAGED AND WORN PARTS HAVE TO BE REPLACED.

Description of Event or Problem · 1

THE FACILITY REPORTS THE BRACKETS FOR THE RAILS AT THE HEAD AND FOOT OF THE TROLLEY WERE BENT AND THAT A PT DROP HAD OCCURRED. SAFETY STAFF WAS INVESTIGATING THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO SHOWER TROLLEY FPO ARJO HOSPITAL EQUIPMENT AB BABXXXX

Patients

Seq Age Sex Outcome Treatment
1