BINAXNOW COVID-19 AG SELF TEST
Report
- Report Number
- 1221359-2021-02075
- Event Type
- Malfunction
- Date Received
- July 20, 2021
- Date of Event
- June 25, 2021
- Report Date
- November 23, 2022
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ADDITIONAL DATA: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 154389 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 154389 , TEST BASE PART NUMBER 195-430H / LOT 151075. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 154389 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.
CORRECTED D3 WITH ACCURATE MANUFACTURER CONTACT INFO. CORRECTED G1 WITH ACCURATE MANUFACTURER CONTACT INFO. CORRECTED G4 TO ACCURATE PMA/510(K) NUMBER EUA210264.
THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.
THE CONSUMER REPORTED 2 UNCONFIRMED FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 AG SELF-TEST PERFORMED ON (B)(6) 2021. THE CONSUMER CONFIRMED THAT HE WAS VACCINATED. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1094489 | BINAXNOW COVID-19 AG SELF TEST | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 154389 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male |