FDA Adverse Event Injury Summary report: N

LOGIC CR TIB INSERT SLOPE+, SZ 4, 9MM

MDR report key: 12192987 · Received July 19, 2021

Report

Report Number
1038671-2021-00347
Event Type
Injury
Date Received
July 19, 2021
Date of Event
June 28, 2021
Report Date
November 22, 2021
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862174482
PMA / PMN Number
K111400
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(H3) AS REPORTED, APPROXIMATELY 4 YRS POSTOP THE INITIAL IMPLANT, THE 83 Y/O FEMALE PATIENT EXPERIENCE KNEE PAIN, AND THE RIGHT KNEE WAS REVISED DUE TO INSTABILITY OF THE TIBIA. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION AS IT WAS DISPOSED BY THE HOSPITAL. UPON REVIEW OF ALL AVAILABLE INFORMATION, THERE IS NO EVIDENCE TO SUGGEST THAT A DESIGN OR MANUFACTURING ISSUE CAUSED OR CONTRIBUTED TO THIS EVENT. THE REPORTED REVISION WAS LIKELY THE RESULT OF UNDERLYING PATIENT FACTORS, WHICH LED TO INSTABILITY OF THE JOINT.

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: LGC TIBIAL FIT TRAY CEM SZ 4F / 4T (CAT# 02-012-45-4040 / SERIAL# (B)(4)), LOGIC CR FEMORAL POR, RIGHT, SZ 4 (CAT# 02-010-04-0340 / SERIAL# (B)(4)), THREE PEG PATELLA 35MM (CAT# 200-02-35 / SERIAL# (B)(4)). ADDITIONAL INFORMATION, INCLUDING THE PRODUCT INVESTIGATION, WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 1

AS REPORTED, APPROXIMATELY 4 YRS POSTOP THE INITIAL IMPLANT, THE (B)(6) Y/O FEMALE PATIENT EXPERIENCE KNEE PAIN, AND THE RIGHT KNEE WAS REVISED DUE TO INSTABILITY OF THE TIBIA. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION AS IT WAS DISPOSED BY THE HOSPITAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1084042 LOGIC CR TIB INSERT SLOPE+, SZ 4, 9MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED JWH EXACTECH, INC. LOGIC CR TIB INSERT SLOPE+, SZ 4, 9MM UNK 10885862174482

Patients

Seq Age Sex Outcome Treatment
1 83 YR Female Required Intervention