FDA Adverse Event
Malfunction
Summary report: N
RICHMAR
MDR report key: 12182363
·
Received July 16, 2021
Report
- Report Number
- 3012316249-2021-00023
- Event Type
- Malfunction
- Date Received
- July 16, 2021
- Report Date
- July 16, 2021
- Manufacturer
- COMPASS HEALTH BRANDS
- Product Code
- IRQ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE HEATING ELEMENT WAS REPLACED 6 WEEKS AGO AND HAVE WORKED FINE SINCE THEN. LAST NIGHT SOMEONE WAS WALKING BY THE AREA AND SMELLED PLASTIC. THE UNIT WAS REPORTEDLY ON FIRE, A FIRE EXTINGUISHER WAS USED, THE ONSITE BIOMED DID A QUICK EVALUATION AND BELIEVES THAT THE FIRE STARTED FROM THE RECENTLY REPLACED HEATING ELEMENT. CONNECTORS TO THE HEATER ARE STILL GOOD. IT DEFINITELY APPEARS THAT THE HEATER (OR WIRES) SHORTED, NOT THE POWER SUPPLY. FUSE DID NOT BLOW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1079503 | RICHMAR | HYDRATHERM DELUXE | IRQ | COMPASS HEALTH BRANDS | HT-R12-D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |