FDA Adverse Event Malfunction Summary report: N

CHLORAPREP ONE STEP HI LITE ORANGE

MDR report key: 12178782 · Received July 15, 2021

Report

Report Number
3004932373-2021-00328
Event Type
Malfunction
Date Received
July 15, 2021
Date of Event
June 21, 2021
Report Date
August 23, 2021
Manufacturer
CAREFUSION 213, LLC 0113
Product Code
KXF
PMA / PMN Number
EXMEPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

FOLLOW UP EMDR PR (B)(4). THE FAILURE MODE OF FOREIGN MATTER (HAIR) WAS CONFIRMED BY THE PHOTOGRAPH PROVIDED BY THE CUSTOMER FOR ANALYSIS. THE HAIR ORIGINATES FROM AN OPERATOR. THE PROCESS HAS CERTAIN MANUAL PROCESSES THAT MAY RESULT IN HAIR ON THE PRODUCT. THE PROCEDURES / PROCESS INCLUDES PREVENTIVE MEASURES FOR HAIR IN PRODUCT. IT IS POSSIBLE FOR ASSOCIATES TO ACCIDENTALLY SHED HAIR ONTO PRODUCT AS THEY ARE LOADING COMPONENTS INTO THEIR CORRESPONDING CONTAINERS /PACKAGES, ALTHOUGH ASSOCIATES WEAR HAIR NETS, GLOVES, SAFETY GLASSES, AND HEAD COVERS, IF WORN IMPROPERLY IT COULD LEAD TO THE REPORTED ISSUE. THE MOST PROBABLE ROOT CAUSE IS INADEQUATE GOWNING BY ASSOCIATE(S) AND/ OR PREVENTIVE MEASURES DURING THE MANUFACTURING OF THE PRODUCT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

MATERIAL NO.: NS930815, BATCH NO.:1063490. IT WAS REPORTED BY THE HEALTH PROFESSIONAL THAT A HAIR AS FOUND INSIDE THE SEALED PACKAGE. PER EMAIL: PLEASE SEE THE ATTACHED PHOTOS. WE HAVE 2 CHLORAPREPS TODAY WITH HAIR INSIDE THE SEALED PACKAGE. THEY ARE FROM THE SAME BATCH, LOT # 1063490, EXP 02/2024. THEY OBVIOUSLY COULD NOT BE USED AS THEY WERE NO LONGER STERILE. I WILL BE PULLING THE REST OF THEM TOMORROW TO DOUBLE CHECK THEM. I HAVE ALSO COPIED OUR LOCAL BD REP ON THE EMAIL.

Additional Manufacturer Narrative · 1

(B)(4). (B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

MATERIAL NO.: NS930815. BATCH NO.:1063490. IT WAS REPORTED BY THE HEALTH PROFESSIONAL THAT A HAIR AS FOUND INSIDE THE SEALED PACKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1074568 CHLORAPREP ONE STEP HI LITE ORANGE 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL KXF CAREFUSION 213, LLC 0113 1063490

Patients

Seq Age Sex Outcome Treatment
1 Other