UPGRADE NEUROWORKS 9.X REVIEW SOFTWARE NON-NATUS PC
Report
- Report Number
- 9612330-2021-00011
- Event Type
- Malfunction
- Date Received
- July 15, 2021
- Date of Event
- July 9, 2021
- Report Date
- October 8, 2021
- Manufacturer
- NATUS MEDICAL INCORPORATED
- Product Code
- GWQ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP REPORT 001 REF NATUS COMPLAINT #(B)(4). PRODUCT EXAMINATION AND FUNCTIONAL TESTING WERE NOT PERFORMED BY NATUS BECAUSE THE CUSTOMER CONFIRMED THE CAUSE OF THE MALFUNCTIONING PART ON-SITE. RISK MANAGEMENT FILE REVIEW: THE CURRENT RISK FILE DOC-010378 - REV 46 HAZARD ID 5.6 IDENTIFIES THE ISSUE. HARM: CLINICALLY INSIGNIFICANT TO DELAYED DIAGNOSIS. A RISK REVIEW IS NOT REQUIRED AS THIS COMPLAINT DOES NOT DESCRIBE A NEW FAILURE MODE OR NEW HARM AND THE EXISTING HAZARD SEVERITY AND/OR PROBABILITY OF OCCURRENCE HAS NOT CHANGED. A SEARCH FOR SERVICE DATA THAT MAY BE RELEVANT TO THIS ISSUE HAS BEEN CONDUCTED. NO FURTHER INFORMATION RELATED TO THIS ISSUE FOUND. INVESTIGATION DETAILS AND INFORMATION ARE STILL UNDER REVIEW.
FOLLOW UP REPORT 002 REF NATUS COMPLAINT # (B)(4). SECTION D4., INCLUDED UNIQUE IDENTIFIER (UDI) (B)(4). PRODUCT EXAMINATION AND FUNCTIONAL TESTING: PRODUCT EXAMINATION AND FUNCTIONAL TESTING WERE NOT PERFORMED BY NATUS BECAUSE THE CUSTOMER CONFIRMED THE CAUSE OF THE MALFUNCTIONING PART ON-SITE. INSTALLATION DATE: (B)(6) 2020. INVESTIGATION RESULTS: (B)(4) WAS CREATED FOR ENGINEERING INVESTIGATION. ISSUE OF PCS FREEZING WAS RESOLVED INTERNALLY WHEN IT RECONFIGURED FIREWALL/PORT BLOCK SETTINGS. INVESTIGATION RESULT CODE: NEURO SBU/SYSTEM CONFIGURATION. CLOSURE RATIONALE: COMPLAINT VERIFIED, ISOLATED INCIDENT AND MONITOR FOR FUTURE OCCURRENCE. NO DEATH OR SERIOUS INJURY, CONSIDERED A CUSTOMER INCONVENIENCE. COMPLAINT WILL BE INCLUDED IN TRENDING DATA FOR FURTHER REVIEW AND INVESTIGATION IF NEEDED.
UPGRADE NEUROWORKS 9.X REVIEW SOFTWARE NON-NATUS PC - A NURSE STATION FROZE AND THE PATIENT EEG WAS NO LONGER BEING MONITORED.
UPGRADE NEUROWORKS 9.X REVIEW SOFTWARE NON-NATUS PC - A NURSE STATION FROZE AND THE PATIENT EEG WAS NO LONGER BEING MONITORED.
INITIAL REPORT REF NATUS COMPLAINT #(B)(4). UPGRADE NEUROWORKS 9.X REVIEW SOFTWARE NON-NATUS PC - A NURSE STATION FROZE AND THE PATIENT EEG WAS NO LONGER BEING MONITORED. A NURSE STATION FROZE AND THE PATIENT EEG WAS NO LONGER BEING MONITORED. THE PATIENT HAD A SEIZURE AND TURNED BLUE. THE PATIENTS' ROOMMATE HEARD THE PATIENT GASPING FOR AIR AND PULLED HIS OWN CALL BUTTON FOR HELP. NO ALERTS TO THE NURSES, NO WAY TO MONITOR OR RESPOND. THE SYSTEM FREEZES UP BETWEEN 6A AND 6:30A , THE LOCAL SUPPORT REP CONTINUES TO ATTEMPT TO FIGURE OUT THE ISSUE. THE ISSUE HAS BEEN ONGOING FOR 1 YEAR. FURTHER INVESTIGATION TO BE CARRIED OUT.
UPGRADE NEUROWORKS 9.X REVIEW SOFTWARE NON-NATUS PC - A NURSE STATION FROZE AND THE PATIENT EEG WAS NO LONGER BEING MONITORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1071654 | UPGRADE NEUROWORKS 9.X REVIEW SOFTWARE NON-NATUS PC | UPGRADE NEUROWORKS 9.X REVIEW SOFTWARE NON-NATUS PC | GWQ | NATUS MEDICAL INCORPORATED | LC103-U9 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |