FDA Adverse Event Malfunction Summary report: N

HLS SET ADVANCED 7.0

MDR report key: 12171590 · Received July 15, 2021

Report

Report Number
8010762-2021-00395
Event Type
Malfunction
Date Received
July 15, 2021
Date of Event
June 30, 2021
Report Date
September 30, 2021
Manufacturer
MAQUET CARDIOPULMONARY GMBH
Product Code
DTZ
PMA / PMN Number
K112360
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT THE DELTA PRESSURE OF A HLS MODULE INCREASED. THE DEVICE WAS BEING USED FOR TREATMENT. THE HLS SET WAS NOT AVAILABLE FOR INVESTIGATION. HOWEVER PATIENT DATA WAS REQUESTED AND ANALYSED BY GETINGE MEDICAL AFFAIRS. THE PATIENT WAS INFECTED WITH COVID-19. AS STATED BY THE GETINGE MEDICAL AFFAIRS TEAM AND IN REFERENCE TO AVAILABLE LITERATURE [1] COVID-19 DISEASES CAN BE ASSOCIATED WITH INTRAVASCULAR COAGULATION ACTIVATION, MICROCIRCULATION DISORDERS AND INCREASED RISK OF THROMBOEMBOLISM DESPITE GOOD SYSTEMIC ANTICOAGULATION. THE INCREASED RISK OF THROMBOSIS AND COAGULOPATHY IN ECMO PATIENTS MAY BE THE RESULT OF A COMBINATION OF PROCESSES DRIVEN BY COVID-19 OCCURRING IN SYNERGY WITH THE KNOWN EFFECT(S) OF THE EXTRACORPOREAL CIRCUIT ON THE COAGULATION SYSTEM. REFERENCE: [1] YUSUFF H, ZOCHIOS V, BRODIE D. "THROMBOSIS AND COAGULOPATHY IN COVID-19 PATIENTS REQUIRING EXTRACORPOREAL MEMBRANE OXYGENATION" ASAIO J. 2020;66(8):844-846. DOI:10.1097/MAT.0000000000001208 IN THE ABSENCE OF FURTHER DATA (PATIENT DATA, TECHNICAL INVESTIGATION) AND BASED ON THE COVID-19 INFECTION THE MOST PROBABLE ROOT CAUSE FOR THE REPORTED FAILURE WAS DETERMINED AS CLOT FORMATION IN THE EXTRACORPOREAL CIRCUIT WHICH CAN LEAD TO A REDUCTION AND/OR BLOCKAGE, AND THUS, AN REDUCTION OF THE DIFFUSION PATH LOWERING THE GAS TRANSFER PERFORMANCE. WITH REFERENCE TO THE RISK ASSESSMENT (HLS SET ADVANCED 5.0 / HLS SET ADVANCED 7.0, DMS # 1468452, V26) AND IN CONSULTATION WITH THE GETINGE MANAGER MEDICAL AFFAIRS TEAM THE FOLLOWING EVENTS CAN CONTRIBUTE TO CLOTTING IN THE CIRCUIT: INAPPROPRIATELY LOW, OR NO BLOOD FLOW, IN THE CIRCUIT, IMPROPER HEMOSTASIS, TOO LOW ANTICOAGULATION, TOO LOW ACT LEVEL, EFFECT OF HEPARIN IS TOO LIMITED, PROTAMINE SULFATE ENTERS THE HLS SET, ADMINISTRATION OF CONGEALABLE SUBSTANCES (SUCH AS PLATELETS) OR PRO-COAGULANT MEDICATIONS. (CONSUMPTION) COAGULOPATHY. THE PRODUCTION RECORDS OF THE AFFECTED HLS MODULE (FAUF#102112420, PO#102112420 AND FAUF#102112413, PO#102112413) WERE REVIEWED ON 2021-09-28. ACCORDING TO THE FINAL TEST RESULTS, THE OXYGENATOR WITH THE SERIAL# 1800995 PASSED THE TESTS AS PER SPECIFICATIONS. PRODUCTION RELATED INFLUENCES ARE UNLIKELY TO HAVE CONTRIBUTED TO THE REPORTED FAILURE. BASED ON THIS THE REPORTED FAILURE "DELTA PRESSURE INCREASED" COULD BE CONFIRMED. THE OCCURRENCE RATE WAS CALCULATED FOR THE REPORTED FAILURE AND IT WAS DETERMINED THAT THIS IS NOT A SYSTEMIC ISSUE. THEREFORE, NO REMEDIAL ACTION IS REQUIRED. THE OCCURRENCE RATE RELATED TO THE REPORTED ISSUE IS CURRENTLY BEING MONITORED AS PART OF MAQUET CARDIOPULMONARY¿ S TRENDING PROGRAM AND ADDITIONAL INVESTIGATIONS OR CORRECTIONS WILL BE IMPLEMENTED IN CASE OF ADVERSE TRENDING. H3 OTHER TEXT : 4114.

Description of Event or Problem · 0

COMPLAINT ID: (B)(4).

Additional Manufacturer Narrative · 1

FURTHER PATIENT DATA AND PERFUSION PROTOCOL WERE REQUESTED BUT NOT YET RECEIVED. A FOLLOW-UP EMDR WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE HLS SET USED ON A COVID PATIENT TREATED FOR ARDS HAD A DELTA PRESSURE INCREASE FROM 15 MMHG TO 122 MMHG, WITH DECREASED FLOWS TO LESS THAN 2.0 LPM. ACT AT THAT TIME WAS 268 S. AS THE PATIENT SATURATION DROPPED THE CIRCUIT WAS CHANGED AFTER 2 H OF INITIATION. PATIENT DID NOT SUFFER ANY COMPLICATION DUE TO DESATURATION AND STABILIZED AFTER THE CIRCUIT EXCHANGE. COMPLAINT ID: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1072341 HLS SET ADVANCED 7.0 OXYGENATOR, CARDIOPULMONARY BYPASS DTZ MAQUET CARDIOPULMONARY GMBH BE-HLS 7050 #SHLS SET ADVANCED 7.0

Patients

Seq Age Sex Outcome Treatment
1 60 YR