HELIOS 1800 TRACK MOUNTED LIGHT
Report
- Report Number
- 1017522-2021-00003
- Event Type
- Malfunction
- Date Received
- July 14, 2021
- Date of Event
- June 24, 2021
- Report Date
- July 14, 2021
- Manufacturer
- KAVO DENTAL TECHNOLOGIES, LLC
- Product Code
- EAZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- 003
Narratives
UPON EVALUATION BY THE DISTRIBUTOR IT WAS DETERMINED THE ROLL PINS WERE NOT INSTALLED BY THE DISTRIBUTOR DURING INSTALLATION. THE ROLL PINS WILL PREVENT THE LIGHT FROM UNSCREWING FROM THE TROLLEY AFTER INSTALLATION. THE PELTON & CRANE INSTALLATION INSTRUCTIONS CLEARLY STATES TO PROPERLY INSTALL THE ROLL PINS DURING INSTALLATION OF THE TRACK LIGHT. THE INSTALLATION INSTRUCTIONS ALSO LIST WARNINGS TO ENSURE THE ROLL PINS ARE INSTALLED. PELTON & CRANE SHIPPED THE ROLL PINS TO THE AUTHORIZED PELTON & CRANE DEALER TO PROPERLY INSTALL THE DENTAL LIGHT PER THE INSTALLATION INSTRUCTIONS. THIS CONCLUDES THE INVESTIGATION.
DEALER TECHNICIAN CALLED KAVO DENTAL TECHNOLOGIES, LLC AND STATED THAT A PELTON & CRANE HELIOS TRACK LIGHT FELL DOWN TOWARDS THE FLOOR HITTING THE PATIENT ON THEIR KNEE CAUSING A BRUISE. THERE WERE NO SERIOUS INJURIES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1065800 | HELIOS 1800 TRACK MOUNTED LIGHT | DENTAL LIGHT | EAZ | KAVO DENTAL TECHNOLOGIES, LLC | HLT |