FDA Adverse Event Injury Summary report: N

TPRLC 133 MP TYPE1 PPS HO 12.0

MDR report key: 12167775 · Received July 14, 2021

Report

Report Number
0001825034-2021-02109
Event Type
Injury
Date Received
July 14, 2021
Date of Event
June 21, 2021
Report Date
September 30, 2021
Manufacturer
ZIMMER BIOMET, INC.
Product Code
KWA
PMA / PMN Number
K110400
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED UPDATED: D4; D9; G3; H2; H3; H4; H6. EVALUATION OF THE RETURNED PRODUCT/PHOTOGRAPHS PROVIDED FOR LOT 6912834 CONFIRMED FOREIGN DEBRIS IS PRESENT INSIDE THE STERILE PACKAGING, AND THE STERILE PACKAGING REMAINS SEALED. THE REPORTED EVENT IS CONFIRMED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. THE LIKELY CONDITION OF THE DEVICE WHEN IT LEFT ZIMMER BIOMET IS NON-CONFORMING TO SPECIFICATION. THE ROOT CAUSE OF THE REPORTED EVENT IS THE OPERATOR NOT FOLLOWING THE PROVIDED WORK INSTRUCTIONS DURING MANUFACTURING AS WELL AS PACKAGING TRANSIT DAMAGE AND PACKAGING DESIGN ISSUES. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION ON THE REPORTED EVENT.

Additional Manufacturer Narrative · 1

(B)(4). REPORT SOURCE: (B)(6). PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR THE INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 -2021 -02104, 0001825034 -2021 -02105, 0001825034 -2021 -02106, 0001825034 -2021 -02107, 0001825034 -2021 -02108.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING WAREHOUSE INSPECTION, DEBRIS WAS FOUND IN THE STERILE PACKAGING. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1062052 TPRLC 133 MP TYPE1 PPS HO 12.0 PROSTHESIS, HIP KWA ZIMMER BIOMET, INC. N/A 6912834

Patients

Seq Age Sex Outcome Treatment
1