FDA Adverse Event Malfunction Summary report: N

THERMO FISHER SCIENTIFIC

MDR report key: 1216534 · Received October 30, 2008

Report

Report Number
MW5008817
Event Type
Malfunction
Date Received
October 30, 2008
Date of Event
October 21, 2008
Report Date
October 24, 2008
Manufacturer
THERMO FISHER SCIENTIFIC, HEADQUARTERS
Product Code
KSE
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
AZ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

RECURRING PROBLEMS WITH BLOOD BANK REFRIGERATOR DOOR HANDLES. DOOR HANDLES (LOCK & STRIKE ASSEMBLY) ARE HAVING TO BE REPLACED ON A QUARTERLY BASIS. THE DOOR HANDLES APPEAR TO BE OF POOR DESIGN AND MADE WITH INFERIOR MATERIALS THAT LEAD TO THEIR CONSTANT BREAKAGE. PRIOR TO THEIR COMPLETE FAILURE, THE HANDLES ALSO CAUSE PROBLEMS WITH THE CLOSURE OF THE REFRIGERATORS' DOORS. WHEN THIS OCCURS, THE REFRIGERATORS EXHIBIT INCREASED STORAGE TEMPS AND TEMP ALARMS RELATED TO THE DOOR NOT CLOSING PROPERLY AND SEALING, AND LEAD POTENTIALLY TO THE LOSS OF BLOOD PRODUCTS BY WAY OF THE INCREASED (EXCESSIVE) STORAGE TEMPS. WE FEEL THAT IF THE DOOR HANDLE WAS OF BETTER DESIGN AND MADE OF BETTER (E.G., STRONGER, MORE DURABLE) MATERIALS, THIS PROBLEMS WOULD NOT OCCUR AS OFTEN AS IT IS NOW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMO FISHER SCIENTIFIC BLOOD BANK REFRIGERATOR KSE THERMO FISHER SCIENTIFIC, HEADQUARTERS BPL25
2 THERMO FISHER SCIENTIFIC BLOOD BANK REFRIGERATOR KSE THERMO FISHER SCIENTIFIC, HEADQUARTERS BBR25
3 THERMO FISHER SCIENTIFIC BLOOD BANK REFRIGERATOR KSE THERMO FISHER SCIENTIFIC, HEADQUARTERS BBR55

Patients

Seq Age Sex Outcome Treatment
1 Other